background image

 

 

 

 

 

AEV350, MEV350

 

TWO PART VALVE HOUSING, 

AUTO AND MANUAL EXPULSION  

       Installation and User Guide

 

32210-001 REV A

 

 

Trulife

 

www.trulife.com                                                                                                                                                                                                                                        

 

 

Page 2 of 4  

 

 

 

 

6.

 

After the socket has cooled, sand off excess plastic directly above the forming tool until the top surface of the forming tool 
is exposed.

 

7.

 

Remove the socket from model. Use a sanding cone to carefully open up the hole if necessary so that the inner housing fits 
snug through the hole. 

 

8.

 

Clean the inside and outside of the formed pocket of anything that could prevent a good seal. 

 

 

Final Assembly: 

9.

 

Make sure the large o-ring is seated in the groove of the flange of the inner housing and place components into the recess 
formed inside the socket. 

10.

 

On the outside of the socket, screw on the aluminum outer housing, and firmly hand tighten. 

Tip: To facilitate a tight fit and a good seal, wet the o-ring and the threads of the housing components with water or 
isopropyl alcohol just prior to tightening.     

11.

 

Assemble the valve insert (auto expulsion or manual expulsion) by gently hand tightening clockwise until the o-ring on the 
outside of the valve insert makes contact with the sealing surface inside the valve housing.  

Note: To prevent cross-threading, it is sometimes necessary to initially turn the insert counter-clockwise until the 
threads are aligned properly, and then tighten clockwise. 

12.

 

A valve key is provided to assist removal of the valve insertfor individuals with limited hand strength or dexterity.  

Disassembly and cleaning valve inserts:  

Foreign  debris  such  as  dust,  baby  powder,  and  residue  from  lotions  used  for  donning  the 
socket  can  become  lodged  in  the  sealing  surfaces  of  the  valve  overtime  and  may  cause  a 
decrease in vacuum performance.  

Instruct  patients  to  periodically  wipe  clean  the  outer  o-ring  and  threads  of  the  valve  insert 
and  the  inner  threads  and  sealing  surface  of  the  valve  housing.  Use  water  or  isopropyl 
alcohol. 

To  disassemble  the  valve  insert  for  more  thorough  cleaning,  steps  1-4  pertain  to  both  the 

auto expulsion and the manual expulsion valve inserts:

 

1.

 

Remove the valve insert from the housing.  

2.

 

Firmly  press  down  on  the  valve  button  until  the  valve  cover 
snaps off. 

3.

 

Maintain  downward  force  of  the  button,  while  carefully 
removing the inner o-ring.  

 

This is made easier by using a small nut or other object on a 
flat, sturdy work surface.  

 

Press the assembly down onto the small object so that the 
valve button lifts up from the inner seal surface. 

 

While  the  valve  button  is  lifted,  use  a  pick  to 

CAREFULLY

 

remove  the  o-ring  from  the  button. 

DO  NOT  CUT  OR 

DAMAGE THE O-RING! 

4.

 

Remove  the  button  and  spring  from  the  valve  body.  Clean 
components  with  water  and  mild  detergent,  or  isopropyl 
alcohol.  Ensure  all  sealing  surfaces  are  free  of  debris  or 
residue. 

Отзывы: