________________________________________________________________________
6
3 TEST FLUID FOR WATER TESTS
To preserve the test liquid for the water tests, mix 45 ml Neoagrar with 15 l
Water. This mixture should have a conductivity between 5.5 and 6.5mS.
4 SETTING THE LC T-T IN STABLE
The LC T-T can be hung with hanging brackets or plug-in brackets or
positioned in the stable.
The devices should hang as vertically as possible.
The hose guide is important. The hoses should be as short as possible. The
Milk must always be able to drain completely. Standing milk would distort
the result of the measurement of the LC T-T.
5 One-time setting
The following settings must be made once.
5.1
Turn on LC T-T
5.2
Put in the USB stick on the LC T-T
5.3
In the main menu, press the button "9" twice in quick succession.
It
appears the password entry for the hidden service menu.
5.4
Enter the six digits password (complementary value 9) to the device number
e.g.:
LC T-T Device number: 070086
Password:
929913
5.5
Confirm with the "ENTER" button.
5.6
Press the "3" button (settings).
Activate measurement online data
5.7
Press button "1" (Online b. Measurement). The selection appears
if you want to measure with or without online data.
In the first line appears the selected option.
5.8
Press button "2" (with online data).
Activate cleaning online data
5.9
Press button "2" (Online b. Clean)
5.10 Press button "2" (with online data) or button “1” (without online data)
Setting language
5.11 Press button "4" (select language)
5.12 Press button "1" (German) » Select your language
5.13 Press the "CLEAR" button, Back to the settings menu
5.14 Switch to the next menu with the "ENTER" button