Appendix A-2
Troubleshooting
GAUGE WILL NOT COMMUNICATE WITH PRINTER OR
COMPUTER
1.
Inspect the contents of the shipping case. If the shipping
case or any other part or accessory appears damaged,
notify the carrier and your Troxler representative
immediately.
2.
Check that the correct cable is being used.
3.
Ensure that the printer/computer baud rate is 9,600 bps.
4.
Ensure that all other parameters match:
♦
Data bits = 8
♦
Stop bits = 2
♦
Parity = none
♦
Protocol = DSR/DTR
GAUGE TURNS OFF AFTER IT IS TURNED ON
1.
To conserve power, the gauge provides an automatic
shutdown if no key is pressed for 60 minutes. To turn the
gauge on, press the power switch.
2.
If the rechargeable NiMH battery pack is below 5.5 V dc,
the gauge shuts down. Recharge or replace the batteries
(see page A–5).