Alarm Settings
The Air XD offers two, user configurable alarms. Each alarm requires four parameters to be selected
and configured to determine the alarm functionality. All alarms are logged in the Air XD event log.
•
Assigned a PM Size/Channel
(PM1.0, PM2.5, PM4.25, PM10.0 and TSP)
Select one of the four configured PM sizes or TSP as the alarm source.
•
Assigned a Data Source/Channel
(Live, Average 1 and Average 2)
Select whether to act on live data readings or one of the pre-configured averages (15min or 8hr).
•
Assigned a Threshold
(µg/m³ and mg/m
3
)
Set the value at which the alarm will trigger (1 µg/m³ to 9999 µg/m³ or 0.1 mg/m
3
to 6553.5 mg/m
3
).
•
Assigned Latching of Non-Latching behavior
Choose whether the alarm stays active once the fault condition is cleared or is self-clearing (non-
latching). For non-latching selections, alarms will be cleared but the event will be logged.
Note:
Once the alarm is triggered, the corresponding relay will then switch, and an alarm message is
displayed on screen.