FLAMMABLE GAS SENSOR/TRANSMITTER
A T E X
M1
GROUP I & II
INTRINSICALLY
SAFE
9 MAINTENANCE
continued
9.4
Annual Safety Check
The main transmitter itself will not normally require maintenance or
calibration but it is advisable to return it to the Trolex Product
Support Department for an annual safety check.
9.5
Damaged Sensors
A Sensor that has been dropped or damaged in any way should
be taken out of service immediately for inspection, repair and
re-calibration.
9.6
Record Keeping
Institute a regular calibration and maintenance procedure and
keep a record.
Incorrect use of the Sensor or inadequate maintenance may not
necessarily be self evident in the Sensor and consequently
it must be regularly checked and maintained.
9.7
Remove Gas Sensing Modules (TX6383.84)
Insert a bar into one of the radial holes in the gas inlet
bush (take care not to damage the internal filter) and unscrew
the gas inlet bush.
Remove the filter from the gas inlet bush. Clean or replace the
filter as necessary. Fit a new or cleaned filter to the gas inlet
bush.
Using a spanner, remove the sensor retaining nut from the
remote housing. Remove the gas sensing module from the
remote housing. Fit a new gas sensing module to the remote
housing, refit the sensor retaining nut and tighten.
Refit the gas inlet bush and tighten using a bar inserted into
one of the radial holes, taking care not to damage the filter.