Gas Inlet Port

 

•  Check the condition and clean with a 

 

 

 

 

 

 

  small brush to remove dust and debris

 

IP ratings do not imply that the equipment will detect gas, during and after  

 

 

exposure to those conditions

 

If it is suspected that the gas inlet port has been exposed to water in    

 

 

conditions that exceed IPx4, the sensor should be dried and the response of    

 

the sensor checked.

Annual Safety Check

The main sensor unit will not normally require maintenance or calibration, but it is 

advisable to return it to the Trolex Product Support Department for an annual safety 

check.

Calibration

The Sentro 1 performs an important safety function and periodic calibration is an 

essential part of maintaining safety integrity.

Gas sensors have a known ZERO and SPAN movement related to time, level of exposure 

to gas, and the nature of the environment, so Trolex recommend that the sensors should 

be bump tested with a suitable test gas at regular intervals. This should be carried out in 

accordance with best practice for the industry where the gas sensor is being used. 

Periodic calibration of the gas sensor should be carried out whilst it is in service. For 

oxygen and carbon monoxide gas sensors Trolex recommends that this is carried out 

every 3 weeks. For other gas sensors Trolex recommends that this is carried out in 

accordance with best practice for the industry where the gas sensor is being used, and 

should take into consideration local operating conditions.

Recommended Test Gas Concentration

Gas Sensor  Type

Zero Gas

Calibration Gas/ balance gas

CH

4

 (Pellistor)  100% LEL

Zero Air

2.2% v/v CH

in air

CH

(Infra Red)

Zero Air or Nitrogen 100%

50% to 100% of range CH

4

 in N

2

CO

Zero Air

50% to 100% of range CO in air

H

2

S    

Zero Air

50% to 100% of range H

2

S in air

NO

Zero Air

50% to 100% of range NO in air

NO

2

Zero Air

50% to 100% of range NO

2

 in air

SO

2

Zero Air

50% to 100% of range SO

2

 in air

CO

2

Nitrogen 100%

50% to 100% of range CO

2

 in air

O

2

Nitrogen 100%

Clean air, or 50% to 100% of range 

O

2

 in N

2

Standard test gas canisters are available from our Product Support department and can 

be supplied in a range of capacities from 34 litres up to 110 litres.
Please call our sales engineers for advice regarding recommended test gas procedures 

and product support plans.
Test gas equipment may be ordered from Trolex using the order code: P5393.23. The kit 

is comprised of a 0.5 litre/min gas regulator, pipe-work and gassing hoods.

Checkpoint

•  CH

4

 test gas must be combined with balance air if being used to calibrate pellistor 

based devices.

•  25% O2 cannot be supplied. A maximum concentration of 23% can be supplied.

www.trolex.com

www.trolex.com

Содержание Sentro 1

Страница 1: ...www trolex com TX6351 TX6352 Sentro 1 Universal Gas Detector...

Страница 2: ...Output Signal 26 3 4 2 4 to 20 mA Analogue Output Signal 27 3 4 3 5 to 15 Hz Output Signal 28 3 4 4 Contact Output Signal 28 3 4 5 RS485 Data Output Signal 29 3 5 Connecting in Hazardous Areas 30 4 S...

Страница 3: ...and TWA monitoring for selected gases 1 2 Application Fixed point gas detection for safety monitoring in hazardous areas and general purpose applications Underground Mining and Tunnelling Ex ia I Ma T...

Страница 4: ...imum Information display 128 x 64 dot graphic backlit LCD Vibration limits Vibration limits IEC 60079 29 1 10 to 30 Hz 1 00 mm total excursion 31 to 150 Hz 19 6 m s2 acceleration peak Impact limits 20...

Страница 5: ...ts RS485 Modbus datacomms 1 6 Electrical Details General Purpose TX6351 3 4 wire Remote powered 4 to 20 mA analogue output Supply voltage 18 to 28 V Max line load 500 R at 24 V Type of sensor Toxic Fl...

Страница 6: ...2 V 20 mA 60 mA 80 mA TX6352 2 wire Remote powered 4 to 20 mA analogue output Supply voltage 10 to 14 V Max line load 90 R Type of sensor Toxic only 1 7 Sentro eModules Plug in pre calibrated modules...

Страница 7: ...ne powered version of Sentro Checkpoint The pellistor is automatically protected against exposure to excessively high concentrations of gas Section 4 4 5 Toxic gases Electrochemical cells Selected tox...

Страница 8: ...used Checkpoint All values listed are nominal and slight variations may occur depending upon operating conditions The natural level of oxygen available in the atmosphere is influenced by relative hum...

Страница 9: ...be calibrated for low gas concentrations and high concentrations up to 100 by volume TX6350 243 245 242 278 2 278 279 Methane CH4 Methane CH4 Methane CH4 Carbon Dioxide CO2 Carbon Dioxide CO2 Carbon...

Страница 10: ...ection in the case of the infrared eModule The Ex certificates associated with the Sentro Modules are listed below eModule Flammable Group I SIRA 10ATEX2046U eModule Toxic Group I SIRA 08ATEX2097U eMo...

Страница 11: ...51 01i 11 and TX9081 01i 11 5 WRT 6 1 WRT 2 or 3 14 4 V 40 mA 135 mW 4 292 mH TX6351 01i 12 and TX9081 01i 12 5 WRT 6 1 WRT 2 or 3 14 4 V 477 mA 1 72 W 4 2 1 mH TX6351 01i 15 and TX9081 01i 15 5 WRT 6...

Страница 12: ...can be affected by other interfering gases which may displace the subject gas being monitored Steam laden atmospheres and condensation can also reduce the sensitivity Flammable Be aware that some toxi...

Страница 13: ...anual to maintain safety reliability and to preserve safety integrity where applicable It is important that sensors are located in positions determined in consultation with those who have specialised...

Страница 14: ...unacceptable voltage error onto the signal This effect can be reduced on long distance connections by increasing the size of the cable cores or even better by running a separate 0 V conductor to power...

Страница 15: ...patible for use with other intrinsically safe circuits emanating from different IS power sources Section 4 4 3 3 4 5 RS485 Data Output Signal RS485 serial data output with analogue signal intelligence...

Страница 16: ...d with approved monitoring apparatus Power for the sensor must be derived from an approved 12 V dc power supply e g TX6642 Intrinsically Safe Power Supply or approved apparatus The installation must b...

Страница 17: ...rev B P5536 1600 rev B 33 4 2 Main Display Power On NEXT FUNCTION SELECT CHANGE Checkpoint Next Function is the Left hand key Left hand key Left L and Select Change is the Right hand key Right hand ke...

Страница 18: ...nu Prompt Bar Calibrate Sentro 1 Setup Output Setup Modbus Setup Module Setup Checkpoint Sentro will automatically return to the Main Display if no commands are given within 1 minute The time limit is...

Страница 19: ...alibrate Zero Follow the prompt bar and screen instructions Go to Span Follow the prompt bar Section 5 2 Calibrate Span using a Test Gas of EXPECTED value Follow the prompt bar and screen instructions...

Страница 20: ...ntinue with the normal span calibration previous page Follow the prompt bar 4 4 2 Sentro 1 Setup The basic functional characteristics of the sensor can be setup or reviewed from a choice of parameters...

Страница 21: ...tup to preferred values Output Information This is the factory setting of the output signal Confirm acceptance Trim Output Zero The level of the transmitted output signal when the sensor is measuring...

Страница 22: ...elay can be independently setup Checkpoint The general relay function will be reversed when monitoring oxygen deficiency i e the relay will de energise when the measured gas concentration recedes from...

Страница 23: ...Change Setpoint 2 The Sentro Module provides two adjustable setpoint output signals When setpoint 1 General is activated the LED alarm indicators in the display window flash simultaneously at a rate o...

Страница 24: ...If the visual alarm has been enabled then this will also activate at level 1 when a STEL or TWA limit is exceeded For a contact output version the STEL and TWA alert will also initiate the Activate c...

Страница 25: ...e by removing it for 2 seconds The gas sensing concentration Module Not Fitted The eModule has been removed and is out for more than 10 seconds Checkpoint Under range output signal will be transmitted...

Страница 26: ...s that this is carried out every 3 weeks For other gas sensors Trolex recommends that this is carried out in accordance with best practice for the industry where the gas sensor is being used and shoul...

Страница 27: ...evaluation and testing of the products with respect to the relevant specific application or use Trolex shall not be responsible or liable for misuse of the information contained herein If you have any...

Страница 28: ...trademarks in this document is acknowledged Document History Revision Date ECN 08 20 11 2014 N A 09 20 01 2015 N A A 11 12 2015 4262 B 25 04 2016 4340 P5536 1600 rev A 25 04 2017 4483 P5536 1600 rev...

Страница 29: ...www trolex com Trolex Ltd Newby Road Hazel Grove Stockport Cheshire SK7 5DY UK t 44 0 161 483 1435 e Sales trolex com www trolex com...

Отзывы: