30
2. Fluid/drug in the actual infusion
line has high viscosity, but the
system occlusion level is set too
low
Raise the alarm Level, Press
【
Start
】
button to continue
infusion
3. The pressure sensor is damaged
Please contact Triumph Medical
Battery nearly
empty
Low
1. When power is supplied only
with the built-in battery, under low
battery,
the
alarm
duration
is >30min
The
alarm
automatically
eliminates after connecting the
external power supply.
2. Battery ageing or the equipment
charging circuit is fault.
Please
contact
Triumph
Medical.
Battery empty High
When power is supplied by the
built-in battery only, under low
battery,
the
alarm
duration
is >30min
Immediately
connect
with
external power supply.
2. Battery ageing or the equipment
charging circuit is faulty.
Please contact Triumph Medical
No battery
inserted
Low
Battery is removed
Keep connected to external
power supply or reinstall the
battery
No power
supply
Low
Under ON state, AC power supply
is adopted, but the AC power wire
is dropped during the process
The
alarm
automatically
eliminates after connecting the
external power supply.
No
batterry
and No power
supply
High
Battery is removed and the AC
power wire is dropped
reinstall the battery or connect
the power supply
Check
syringe
High
Syringe drop off during infusion
Reinstall the syringe
Reminder
alarm
Low
After installing syringe tube, under
non-working or alarm state, it is
not operated within the set time of
the system
Click any button to stop
Standby time
expired
Middle
During standby, after reaching the
standby time
Press
【
Stop
】
button to stop
alarm
KVO finished
High
KVO working time reaches 30min,
syringe pump stops working
Press
【
Stop
】
button to stop
alarm
Содержание SP12
Страница 1: ...User Manual Triumph SP12 Syringe Pump ...
Страница 15: ...14 ...
Страница 33: ...32 ...