Measurement Calibration
- The calibration can be adjusted
in two ways by using the slider control to make adjustments.
a.
Calibration by Measurement
Place the probe onto a test piece of known thickness
and adjust the slider control until the correct reading is
achieved. The
Velocity of Sound
display will show the
calibrated value for this sample.
b.
Calibration by Velocity
When a suitable test sample is unavailable, calibrate
the gauge by adjusting the
Velocity of Sound
to an
appropriate value (see the velocity table in the gauge
manual).
8.3.6.
Diagnostics Tab
Diagnostics should not normally be required when using Tritex gauges, the
information is provided to check the correct performance of the gauge and link.
The only control provided on the
Diagnostics tab
is the
PING
button. Click on
PING
at any time and the gauge should respond immediately. The contents of
the text box is cleared and replaced with information identifying the connected
gauge type.