2
Important Safety Instructions
Statement of Intended Use
Tripp Lite Medical-Grade UPS Systems are intended to support and protect non-medical computer
equipment and medical devices that require leakage current reduction, surge protection, voltage
regulation, line noise filtering and battery backup during power outages and generator testing, both
inside and outside patient care areas. Tripp Lite’s Medical-Grade UPS Systems come with hospital-
grade plugs and receptacles that reduce leakage to below 100μA.
Product Information
OMNIX350HG
SMX700HG
SMX1200XLHG
AC Input Voltage
230V
230V
230V
Input Amp(s)
1.78
4
5
Output Amp(s)
1.5
2
3.3
Rated Power (VA/Watts)
350VA, 225W
700VA, 450W
1000VA, 750W
Frequency
50/60 Hz
50/60 Hz
50/60 Hz
No. of Phases
Single
Single
Single
Class Type
Class I
Class I
Class I
Plug/Connector Type
Inlet C14 with dongle Inlet C14 with dongle Inlet C14 with dongle
Operation
Continuous
Continuous
Continuous
AC Input Protection (Qty/type/rating)
Input breaker (2 x 3A)
thermal
Input breaker (2 x 5A)
thermal
Input breaker (2 x 6A)
thermal
Max Leakage Current (uA)
Less than 100uA
Less than 100uA
Less than 100uA
Weight
11.5 kg
19.1 kg
20.6 kg
Known Contraindication
None
None
None
SAVE THESE INSTRUCTIONS
This manual contains important instructions that should be followed during the installation,
operation and storage of all Tripp Lite UPS Systems. Failure to heed these warnings may affect
your warranty.
Note:
Your UPS incorporates overcurrent protection in both the L1 and L2 input conductors.
UPS Location Warnings
Relevant Symbols
Denotes that information in the manual should be reviewed before use.
Used to signal as a warning that a statement is particularly important and could pose a safety
risk.
Warning-Dangerous Voltage
No sitting
Denotes general warning sign
No stepping on surface
Refer to Instruction Manual/Booklet
No pushing