Triplett MS475 Скачать руководство пользователя страница 6

Moisture Alarm Set Procedure: 

 

9. 

With MOIST displayed, simultaneously press the REL button. 

 

10. 

The “HIGH” icon will appear on the display. 

 

11. 

Press the ▲or ▼button to set the high limit desired. 

 

12.  Press the REL button to save the value and proceed to the LOW set value. 

 

13. 

With the “LOW” icon in the display, Press the ▲or ▼button to set the low limit desired. 

 

14.  Press the STORE/ALARM SET button to save the value and to return to the normal mode. 

 

15.  If the moisture measurement is higher than the LOW alarm setting, the meter will beep once 

every second 

16.  If the moisture measurement is higher than the HIGH alarm setting, the meter will beep 

continuously. 

 

Infrared Temperature Measurements 

 

1. 

Press the power 

 button to turn the meter on. 

 

2. 

Press the IRT button to enable the IR thermometer and the laser pointer. The laser pointer 
icon will flash while the mode is active. 

 

3. 

Aim the laser pointer at the surface to be measured and read the surface temperature in the 
secondary display. 

 

4. 

Release the IRT button. The last temperature measured and the laser icon will remain on the 
display for approximately 10 seconds before returning to ambient temperature measurement. 

 

IRT MAX MIN display: 

 

The meter can be set to display only the maximum or minimum temperature measured during an IR 
scan. 

 

1. 

With the meter in the IR hold mode, press the 

▼/zero button. “MIN” will appear in 

the display. 

 

2. 

Press the IRT button to enable the IR thermometer. The meter will display the minimum 

temperature measured and will update only when a lower temperature is measured. 

 

3. 

Press the 

▼/zero button twice to enable the MAX mode and proceed as stated above for 

the maximum temperature. 

 

4. 

The MAX or MIN temperature is not stored when the function is exited. The unit 
automatically exits MAX/MIN mode after approximately 10 seconds. 

 

IR Field of View 

 

Ensure that the desired target is larger than the spot size. As the distance from an object increases, 
the spot size of the area measured by the meter becomes larger. The meter’s field of view ratio is 
8:1, meaning that if the meter is 8 inches (cm) from the target, the diameter (spot) of the object 
under test must be at least 1 inch (cm). 

 
 

WARNING: Do not directly view or direct the laser pointer at an eye. 

Low power visible lasers

 

do not normally present a hazard, but may present some potential for hazard if viewed directly for 
extended periods of time. 

Содержание MS475

Страница 1: ...lus IR Thermometer Monitor moisture in wood and other building materials with no surface damage with the Pinless Moisture sensor Pin type Moisture Probe included Measure Humidity and Air Temperature w...

Страница 2: ...COND Condensation Button 5 RH Relative Humidity Button 6 MOIST REL Button 7 DOWN ZERO Button 8 UP UNIT Button 9 Memory Store Enter ALARM Set Button 10 IR Thermometer Button 11 ON OFF Button 12 Pinles...

Страница 3: ...tion 6 Low battery indication 7 GPP unit indication Grains Per Pound 8 MAX MIN indication 9 Condensation mode symbol 10 VPP unit indication Vapor Pressure 11 GPP unit indication 12 Humidity bar displa...

Страница 4: ...ure plus Grains Per Pound GPP and Dew Point DP Ambient and Surface Vapor Pressure Automatic calculation of differential Temperature IR DP Min Max and Data Hold 20 point internal memory Auto power off...

Страница 5: ...e and to return to the normal mode 6 If the humidity measurement is higher than the LOW alarm setting the meter will beep once every second 7 If the humidity measurement is higher than the HIGH alarm...

Страница 6: ...r approximately 10 seconds before returning to ambient temperature measurement IRT MAX MIN display The meter can be set to display only the maximum or minimum temperature measured during an IR scan 1...

Страница 7: ...ew Point the temperature difference shall be displayed with no other warning If the temperature of the IRT is 3 14 C 5 25 F above the Dew Point the temperature difference shall be displayed along with...

Страница 8: ...ay stored readings The numeric display above the MEM icon will flash 2 Use the or button to scroll through the memory locations 3 To return to normal operating mode press the STORE button Clearing Sto...

Страница 9: ...to 1 C 9 F 5 C D S 8 1 32 F 0 C 1 8 F 1 C Emissivity 0 95 Fixed 33 to 392 F 1 to 200 C 3 6 F 2 2 C Display 3 digit primary display 4 digit secondary display Vapor Pressure 0 to 20 0kPA calculated fro...

Страница 10: ...hese goods for use Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of 1 one year from the date of pu...

Отзывы: