D04-05Aen201908 Quick Guide TTrig

TriOS Mess- und Datentechnik GmbH · Bürgermeister-Brötje-Str. 25 · D-26180 Rastede · Germany 

Phone: +49 (0) 4402 69670 - 0 · Fax: +49 (0) 4402 69670 - 20 · [email protected] · www.trios.de

In comment line 3 each measurement is assigned a consecutive number.
Failed measurements, e.g. due to service or power fail, are internally incremented in TTrig. 
In this case, the number of the subsequent measurement is not consecutive.
The numbering of the measurements enables the user to subsequently correct fluctuations of the sensor-internal RTC (Real Time 

Clock).

Comment line 4 contains additional event information suchas timeout, power fail, etc.
In the case of omitted measurements, their number can be read at "EVENT-SKIPPED #".

TTrig Reset

Reset resets the TTrig to its initial position. 

Automatic measurement is stopped and the numbering of the measurements is reset to 1.

1. INTERVAL and 2. DELAY can now be reconfigured. By pressing the start button the TTrig takes over the settings of the rotary 
encoders into the new automatic measurement.

All stored data of the sensor is retained!

If the web interface is also not available, please contact us at [email protected].

Troubleshooting

Power fail

A power fail occurs when the power supply is no longer sufficient for the next measurement. The red POWER FAIL LED flashes and 
no more measurement is triggered in the following 12 hours. Every 12 hours, the TTrig will automatically check the performance 
of the power supply.

The start button can be used to trigger a manual check of the voltage source.

The TTrig resumes automatic measurement when sufficient power is available.

Sensor not found:

Check the connector and the cable of the sensor. Carry out a reset and press the start button again after one minute. 

Sensor detected 

  Measurements are carried out fully automatically.

Sensor not detected 

 Connect to the G2 interface of the sensor and check the settings of the serial interface under the menu 

"Peripherals". 

Содержание TTrig 12C100000

Страница 1: ... Remote or self sufficient measuring stations can thus be operated maintenance free for several months An RJ 45 Ethernet interface provides access to the sensor s G2 web interface for downloading the measurement data from the data logger with a notebook Technical Specifications POWER SUPPLY Voltage supply 12 24 VDC max 4A Power in standby 1mW SENSOR INTERFACES Connection M12 for TriOS G2 sensors 1...

Страница 2: ...permanently for data download Sensor single measurement Wiper activation Reset reset program If the red power fail LED flashes this button can be pressed to start a manual attempt to restore normal operation LEDs Description Colour Function Behaviour Start white Measurement interval active LED flashes briefly once every 20 seconds white Sensor initialisation LED flashes continuously Wiper white Wi...

Страница 3: ...nfiguration options Function test The WIPER and SENSOR buttons should be operated manually one after the other for the subsequent function test The function of the wiper is visible by its movement For OPUS it is recommended to download the data logger and check the completeness of the measurements Configuration of automatic measurement The measurement interval is set on site on the TTrig rotary en...

Страница 4: ...ice LED flashes every 20 s please refer to the section Troubleshooting Sensor not found Data storage and retrieval All measurement data is only stored on the internal 2GB SD card of the sensor A notebook can be connected via an RJ 45 socket on the TTrig so that the stored measurement data can be downloaded via the web interface Press the Service button on the TTrig to supply the sensor with power ...

Страница 5: ...set to 1 1 INTERVAL and 2 DELAY can now be reconfigured By pressing the start button the TTrig takes over the settings of the rotary encoders into the new automatic measurement All stored data of the sensor is retained If the web interface is also not available please contact us at support trios de Troubleshooting Power fail A power fail occurs when the power supply is no longer sufficient for the...

Отзывы: