D01-905en201912 Manual eCHEM Free Chlorine
Free Chlorine
//
Malfunction and Maintenance
6.3.1 General Troubleshooting
Fault
Possible cause
Action
Sensor cannot be calibrated / measured
value deviates from DPD measurement
Gas bubbles in the electrolyte
Unscrew the membrane cap and empty
the electrolyte.
Repeat commissioning.
Run-in time too short
Repeat calibration after a few hours
Membrane cracked
Replace membrane cap (see chapter
6.2.2)
Membrane cap damaged
Replace membrane cap (see chapter
6.2.2)
Interfering substances in the water
Examine water for interfering substan-
ces and remedy, consult the supplier if
necessary
Short-circuit / defect in the measuring
line
Locate and eliminate short-circuit / de-
fect, replace measuring line if necessary
Distance between membrane and
electrode is too great
Screw on the membrane cap complete-
ly as far as it will go
DPD chemicals too old
Use new DPD chemicals and repeat
calibration
Deposits on the membrane
Replace membrane cap (see chapter
6.2.2)
Gas bubbles on the outside of the
membrane
Briefly increase the flow rate, check and
change the installation if necessary
Sensor defective
Send in sensor to be checked / over-
hauled
Missing galvanic isolation
Establish galvanic isolation.
Send in sensor to the supplier to be
checked / overhauled.
No electrolyte in the membrane cap
Fill the membrane cap with electrolyte
(see chapter 6.2.1)
Chlorine content exceeds upper measu-
ring range limit
Check system, remedy fault, repeat
calibration
Unstable measuring signal
Membrane cracked
Replace membrane cap (see chapter
6.2.2)
Bubbles in the electrolyte (especially
during operation under pressure)
Empty the membrane cap and fill
with new electrolyte bubble free (see
chapter 6.2.1)
Gas bubbles on the outside of the
membrane
Briefly increase the flow rate, check and
change the installation if necessary
Pressure fluctuations in the measuring
medium
Check installation type and change if
necessary
Reference electrode exhausted and/or
contaminated
Send in sensor to be overhauled
Missing galvanic isolation
Establish galvanic isolation.
Send in sensor to the supplier to be
checked / overhauled.