27
D01-903en201912 Manual Dissolved Oxygen
Dissolved Oxygen
//
Annex
These setting are only saved temporarily and are lost after a restart.
Unit
Register
00 00 00 00 0_ _ _ 00 00
%
166
mg/L
167
ppm
168
Salinity (0=inactive oder 1=active)
Pressure (0=inactive oder 1=active)
Temperature (0=internal oder 1=external)
The adjustments for each parameter can be activated separately. The value 1 hereby activates the external correction, whereas
0 deactivates it. The temperature is an exception. 0 is used for the internal temperature correction.
To use an external temperature measurement, it has to be written into the registers 93/94 as a float (unit °C) and activated in
the registers 166,167 or 168 with 00 00 00 00 0
0 01
00 00.
To use a correction for the atmospheric pressure, it has to be written into the registers 95/96 as a float (unit hPa) and activated
in the registers 166,167 or 168 with 00 00 00 00 0
0 10
00 00.
To use a correction value for the salinity, it has to be written into the registers 97/98 as a float (unit ppt) and activated in the
registers 166,167 or 168 with 00 00 00 00 0
1 00
00 00.
To activate more than one correction, it is possible to set all Bits to 1 (active).
Example: Sample for register 166, 0167 and 168.
External compensation of temperature, salinity and atmospheric pressure
HEX
DEC
BIN
0x0070
112
00 00 00 00 0
1 11
00 00
All three corrections are activated
0x0060
96
00 00 00 00 0
1 10
00 00
Salinity and pressure activated, temperature internal
0x0040
64
00 00 00 00 0
1 00
00 00
Salinity activated, pressure deactivated and temperature internal
NOTICE