D01-903en201912 Manual Dissolved Oxygen
Dissolved Oxygen
//
Annex
Number
Parameter
Unit
1
Temperature
°C
2
Oxygen saturation
%
3
Oxygen concentration mg/l
4
Oxygen concentration ppm
Modbus RTU
Parameters
Supported Modbus functions
These Modbus function codes are supported by the sensor:
Name
Code
Description / Use
Read multiple registers
0x03
Read the serial number, firmware version and of course measurement
data
Write single register
0x06
Write a value in one register
Write multiple registers
0x10
Write data into a sequence of registers
Report slave ID
0x11
Get the sensor identification
Name
Register Count Format
Uint16
1
Unsigned 16 bit integer. Value range: 0x0000 - 0xFFFF
Bits16
1
Register contains a bitmask, where every bit has a special meaning. This is most
often used for parameter selection, where Bit 1 corresponds to the temperature
and the following bits to parameter 1, 2, 3 and 4
Bits32
2
A bit vector using 32 bits, same as Bits16 but spanning two registers
Uint32
2
Unsigned 32 bit integer. Value range: 0x00000000 - 0xFFFFFFFF
Float
2
IEEE 754 32 bit floating point value
ASCII
1+
A sequence of ascii characters with two 8 bit characters stored in each 16 bit
register
Date
8
A date in the form mmhhddMMYYYY, where mm is the minute, hh the hour, dd
the day, MM the month and YYYY the year
Data types
All multi-register datatypes are stored in big-endian word order. That is the word with the most significant bits is stored at the
lesser register address.