Страница 1: ...Dissolved Oxygen OPERATING INSTRUCTIONS...

Страница 2: ......

Страница 3: ...Installation 9 4 3 Support Tube Installation 10 5 Calibration 12 5 1 Two Point Calibration 12 5 1 1 Offset Calibration 12 5 1 2 Slope Calibration 12 5 2 One Point Calibration 13 5 3 Measurement Prope...

Страница 4: ...sensors and accessories supplied by TriOS Mess und Datentechnik GmbH must be installed and operated in accordance with the specifi cations provided by TriOS Mess und Datentechnik GmbH All parts were...

Страница 5: ...tion may cause moderate injury Electromagnetic Waves Devices that radiate strong electromagnetic waves can influence the measurement data or result in a malfunction of the sen sor Avoid using the foll...

Страница 6: ...lified experts authorized byTriOS Devices from TriOS Mess und Datentechnik GmbH meet the highest safety standards Repairs to the device that involve the replacement of the connecting cable must be don...

Страница 7: ...to use when it is handled according to the instructions in this user manual 1 6 Disposal Information At the end of the device s life or use the device and its accessories can be returned to the manuf...

Страница 8: ...sensor can even be used in applications with very weak flow of water The sensor offers the following advantages Low operating costs due to reduced maintenance work no electrolyte changes Greater calib...

Страница 9: ...Warranty Customer Service Contact Keyword Index Accessories 2 2 Scope of Delivery The shipment contains the following components Sensor Operating Instructions Accessories if applicable Keep the origin...

Страница 10: ...Ensure correct polarity of the operating voltage or otherwise the sensor may be damaged 1 RS 485 A 2 RS 485 B 6 Power 12 VDC 7 Ground Power Ser Interface NOTICE The sensor is ready for initial startup...

Страница 11: ...r water Standard operation of the sensor is carried out by immersing it into the medium The sensor can either hang freely or be mounted at 45 For measurement you must eliminate bubbles trapped under t...

Страница 12: ...00B00 4136 ZM111710 Mechanik Halterohr 9 Version 1 1 Update January 20 AQUALABO CONTROLE 35 Rue Michel Marion 56850 CAUDAN FRANCE Tel 33 0 2 97 89 25 30 Fax 33 0 2 97 76 55 72 Long pole The long poles...

Страница 13: ...unction Maintenance Commissioning Introduction General Information FAQ Technical Data Warranty Customer Service Contact Keyword Index Accessories 2 Pull the complete cable through 3 Fix the sensor wit...

Страница 14: ...ut as follows 5 1 1 Offset calibration The sensor must be cleaned before calibration see chapter 6 Immerse the sensor in a water sulphite solution sulphite con centration 2 in order to determine the z...

Страница 15: ...t electronics in the form of functions The sensor can therefore determine the oxygen concentration of the measurement medium in order to digitally transmit the influencing factors above in compensated...

Страница 16: ...peration it should be rinsed prior to being stored and the protective cap should be fitted with the protective case and a moist absorbent surface like cotton 6 1 Cleaning and Upkeep Rinse meticulously...

Страница 17: ...avoid incorrect deliveries each return package must first be reported to the customer service You will then receive a numbered RMA form which you need to fill out completely check and send back to us...

Страница 18: ...ion Via NTC compensation active for temperature lower than 0 C Measurement range temperature 0 50 C Resolution temperature 0 01 C Accuracy temperature 0 5 C Membrane Cap No cross sensitivity with pH 1...

Страница 19: ...cessories Calibration maintenance interval 2 years Warranty 1 year EU US 2 years on electronics All wearing parts are not included in the warranty INSTALLATION Max pressure 5 bar 72 5 psig Protection...

Страница 20: ...able A built in valve allows the use of compressed air cleaning for the sensors The TriBox3 also has various interfaces including an IEEE 802 3 Ethernet inter face an IEEE 802 11 b g n interface a USB...

Страница 21: ...sumables such as light sources are not included in the warranty The warranty is subject to the following conditions The device and all accessories must be installed as described in the corresponding m...

Страница 22: ...Customer Service If you are having a problem with the sensor please contact TriOS customer service To send back the sensor please request an RMA number from customer service Technical support contacts...

Страница 23: ...ssories 11 Contact We are constantly working to improve our devices Visit our website for news and information If you have found an error or bug in one of the devices or programs please let us know Cu...

Страница 24: ...pressure 13 B Biological safety 3 Bypass installation 9 C Calibration 12 CE certification 24 Certificates approvals 5 Cleaning 14 Contact 21 Copyright 2 Customer service 20 D Disposal 5 E Electrical i...

Страница 25: ...tact Keyword Index Accessories P Product Identification 6 Q R Rating plate 6 Reagents 3 Returns 15 RMA number 15 S Safety instructions 3 Salinity 13 Scope of delivery 7 Slope Calibration 12 Specificat...

Страница 26: ...res conformity of the product TriOS GmbH d clare la conformit du produit Bezeichnung eCHEM Sauerstoff Product name eCHEM Oxygen Designation Typ Type Type Art Nr 90S531100 90S531130 Art Nr 90S534100 90...

Страница 27: ...igned 16 bit integer Value range 0x0000 0xFFFF Bits16 1 Register contains a bitmask where every bit has a special meaning This is most often used for parameter selection where Bit 1 corresponds to the...

Страница 28: ...and 1 stop bit Temporary coefficients W 332 Bits32 Each calibration coefficient Offset Scaling has a corresponding bit in this two registers Bit 1 Offset of temperature sensor Bit 2 Scaling of tempera...

Страница 29: ...8 with 00 00 00 00 00 01 00 00 To use a correction for the atmospheric pressure it has to be written into the registers 95 96 as a float unit hPa and activated in the registers 166 167 or 168 with 00...

Страница 30: ...r averaging perform offset calibration hold sensor in clean water 10 Min No Yes move sensor slowly in a Na2 SO3 solution for 5 minutes 2 Measurement point 1 Measurement point trigger measurement and r...

Страница 31: ...ion Na2SO3 a tip of a teaspoon dissolved in water and move the sensor around in the solution Leave the sensor in the solution until the measurement values have stabilized 5 Measurements have to be tri...

Страница 32: ...nt Point by putting the sensor in clean water for 10 minutes taking the sensor out holding it in air and continuously perform measurements in air and reading the results Saving the calibration 10 To a...