SafeTcheck OPeRaTOR'S MaNUaL
20
Note 1:
Pressing RESET at any point during the test
will shorten the test duration. This does not invalidate
the test result.
Note 2:
The
SafeTcheck
does not use
to test for the presence of a return lead/probe in the Run
Test mode. Failing to use the return lead/probe while
running the appliance invalidates the test.
Note 3:
If Data Logging is switched on then additional
information will be requested. See section entitled
'Data Logging'
Operating Instructions
1 Ensure the appliance is switched off.
2 Plug the appliance into the separate Run Test socket.
4 Connect the return lead/probe to the
SafeTcheck
via the
Earth Return Socket. Use the Probe to wipe over the
surface of the appliance and make contact with metal parts
during the test.
WaRNINg
The appliance will operate for up to 20 seconds in the run
test mode. Ensure rotating or heating elements are able to
operate safely. Appliances such as drills, angle grinders etc
should be firmly restrained during the run test.
5 Switch the appliance on.
6 Select Class 2 Run Test under Other APPLIANCE menu
using the F1 and F2 keys.
7 Press START button and follow the screen instructions. The
appliance will operate for up to 20 seconds during the run test.
WaRNINg
Upon completion of a safety test and subsequent
removal of test lead, please ensure all switches, speed
or thermostat controls are returned to their ‘OFF’ or
minimum position.
Run testing double insulated (class 2) appliances
Note:
Use the probe to make contact with metal
exposed parts and insulated surfaces. See section
entitled "Utilities Menu" for more information about
the Probe Mode