Trio CS-1566A Скачать руководство пользователя страница 14

1

. Apply the type of signal normally encountered 

in the amplifier under test. 

2. Connect CH1 probe to the input of the 

amplifier and CH2 probe to the output of the 

amplifier. It is preferable if the two signals are 

not inverted in relationships to each other, but 

inverted signals can be used. 

3 Set CH1 and CH2 AC-GND-DC switches to 

AC. 

4. Set the MODE switch to ALT or CHOP. 

5. Set SOURCE switch to CH1 and adjust 

controls as described in the procedure for 

synchronizing waveforms. 

6. Adjust CH1 and CH2 POSITION controls to 

superimpose the waveforms directly over each 

other. 

7. Adjust CH1 and CH2 vertical sensitivity con-

trols (VOLTS/DIV and VARIABLE) so that the 

waveforms are as large as possible without ex-

ceeding the limits of the scale, and so that 

both waveforms are exactly the same height. 

8. Now, set the Mode switch to ADD and the 

CH2 polarity switch to the PULL INVERT posi-

tion. (If one waveform is inverted in 

relationships to the other, use the normal CH2 

polarity.) 

Adjust the vertical sensitivity control (CH2 

V A R I A B L E ) for the minimum remaining 

waveform. Any waveform that remains equals 

distortions; if the two waveforms are exactly 

the same amplitude and same waveform and 

there is no distortion, the waveforms will 

cancel and there will be only a straight 

horizontal line will remaining on the screen. 

Gated Ringing Circuit (burst circuit) 

The circuit and waveform of Fig.

 11

 are shown to 

demonstrate the type of circuit in which the 

dual-trace oscilloscope is effective both in design 

and troubleshooting applications. 

Fig.

 11

 shows a burst circuit. The basic settings 

are identical to those of in Fig. 5. 

Waveform EA is the reference waveform and is 

applied to CH1 input. All other waveforms are 

sampled at CH2 and compared to the reference 

waveform of CH

 1.

 The frequency burst signal can 

be examined more closely either by increasing the 

sweep time per division to

  0 . 5

 ms per division or by 

pulling the POSITION control to obtain

 10 

times magnification. This control can then be 

rotated as desire to center the desired waveform 

information on the oscilloscope screen. 

Delay Line Test: 

The dual-trace feature of the oscilloscope can also 

be used to determine the delay times of transmis-

sion type delay lines as well as ultrasonic type 

delay lines. The input pulse can be used to trigger 

or synchronized the  C H

1

 display and the delay line 

output can be observed on CM2. The interval 

between repetitive pulses should be large com-

pared to the delay time to be observed. In addition 

to determining delay time, the pulse distortion in-

herent in the delay line can be determined by ex-

amination of the delay pulse observed on CH2 

waveform display. 

Fig.

 12

 shows the typical oscilloscope settings as 

well as the basic test circuit. Typical input and out-

put waveforms are shown on the oscilloscope dis-

play. Any pulse stretching and ripple can be 

observed and evaluated. The results of modifying 

the input and output terminations can be observed 

directly. 

A common application of the delay line checks is 

found in color television receivers to check the  " Y " 

delay line employed in the video amplifier section. 

The input waveform and output waveform are com-

pared for delay time, using the horizontal sync 

14 

Fig.

 11

 Gated ringing circuit and waveforms 

Содержание CS-1566A

Страница 1: ...TRIGGERED SWEEP OSCILLOSCOPE HIGH S T A B I L I T Y CS 1566A DUAL TRACE OSCILLOSCOPE INSTRUCTION MANUAL T R I O...

Страница 2: ...observation The time base switch allows changeover between V vertical and H horizontal of VIDEO sync separator circuit automatically and electronically INT C H 1 CH2 LINE and EXT can be individually...

Страница 3: ...time 0 5Ms div to 5s div 5 and X Y 1 2 5 sequence Fine adjustment in all 19 ranges Magnification 10 times 5 PULL x 10 MAG Linearity Better than 3 2 zs div to 0 5s div Better than 5 0 5 s div to 1 s d...

Страница 4: ...tation Trace angle is adjustable by panel surface ad justor Power Requirements Power supply voltage 100 120 220 240V 10 50 60 Hz Power consumption 47W Dimensions Width 260 mm 277 mm Height 190 mm 204...

Страница 5: ...CONTROLS ON PANELS FRONT PANEL Fig 1 REAR PANEL Fig 2 5...

Страница 6: ...Dual trace operation in which sweep is chopped at approximate 200 kHz rate and switched between Channel 1 and Channel 2 traces Recommended for sweep times of 0 5 s div to 1 ms div ADD The waveforms f...

Страница 7: ...tive going sync pulse 19 LEVEL PULL AUTO Sync level adjustment determines points on waveform slope where sweep starts equals most negative point of triggering and equals most positive point of trigger...

Страница 8: ...40V be sure to use a 0 5A fuse 29 POWER CONNECTOR For connection of the supplied AC power cord 30 Z AXIS INPUT Intensity modulation terminal T T L logic compatible high logic increase brightness low l...

Страница 9: ...djust VOLTS DIV 4 7 and VARIABLE 5 8 so that the pulse amplitude is set to the 0 and 100 scales 2 Turn SWEEP TIME DIV 21 to magnify the rising portion of the waveform as large as possible VARIABLE 22...

Страница 10: ...ative going waveform In this case the output waveform is shifted with respect to the leading edge of the Fig 5 Waveforms in divide by two circuit reference frequency pulse by a time interval cor respo...

Страница 11: ...of the waveforms change depending upon the input or more mode of opera tion Ffg 8 shows a typical digital circuit and identifies several of the points at which waveform measure ment are appropriate T...

Страница 12: ...uch cases it is recommended that the sync re main unchanged while the sweep speed or X 1 0 Fig 8 Typical digital circuit using several time related waveforms MAG used be to expaned the waveform displa...

Страница 13: ...e amplifier unit may be measured for distortion with this oscilloscope This type of measurement is especially variable when the slope of a waveform must be faithfully reproduced by an amplifier Fig 10...

Страница 14: ...ns Fig 11 shows a burst circuit The basic settings are identical to those of in Fig 5 Waveform EA is the reference waveform and is applied to CH1 input All other waveforms are sampled at CH2 and compa...

Страница 15: ...Y P E T O H O R I Z O N T A L A M P L I F I E R T O V I D E O O U T P U T A M P L I F I E R N O T E NO E L E C T R I C A L C O N N E C T I O N P L A C E C L I P ON I N S U L A T I O N O F P L A T E C...

Страница 16: ...ut jack The VARIABLE control is adjusted as required to provide a complete cycle of the input waveform displayed on 8 div horizontally A waveform height of 2 div is used The 8 div display represents 3...

Страница 17: ...abled to see the VITS The transmitted VITS is precision sequence of a specific frequences amplitudes and waveshapes as shown in Fig 16 Television networks use the precision signals for adjustment and...

Страница 18: ...s on either side This suggests an IF trap is detuned into the passband chopping out frequencies about 2 MHz below the picture carrier frequency Switch to another channel carrying VITS If the same thin...

Страница 19: ...ection moving the trace to the left until the expanded VITS information appears as shown in Fig 16 Because of the low repetition rate and the high sweep speed combination the brightness level of the s...

Страница 20: ...he peak to peak voltage readings on the oscilloscope with those shown on the waveform pictures Any abnormal readings should be followed by additional readings in the suspected circuits until the troub...

Страница 21: ...controls for the desired brightness and best focus 15 To view a specific portion of the waveform such as the color burst pull outward on the POSITION control for X10 magnification Rotate the same con...

Страница 22: ...tal gain controls for a display similar to that shown in Fig 23A 5 Set the marker generator precisely to 10 7 MHz The marker pip should be in the center of the bandpass 6 Align the IF amplifiers accor...

Страница 23: ...e waveform A sine wave input is applied to the audio circuit being tested The same sine wave input is applied to the vertical input of the oscilloscope and the output of the tested circuit is applied...

Страница 24: ...ces a circular oscilloscope pattern Phase shift of less or more than 90 produces an elliptical oscilloscope pattern The amount of phase shift can be calculated from the oscilloscope trace as shown in...

Страница 25: ...can evaluate the input and output quality of a signal of many frequencies the harmoniques of the square wave which is what the amplifier sees when amplifying complex wave forms of musical instruments...

Страница 26: ...ase a deliberate non linear circuit such as a clipper network 3 The third is delay or phase distortion which is distortion produced by a shift in phase between one or more components of a complex wave...

Страница 27: ...lts are opposite in the two cases because of the difference in polarity of the phase angle in the two cases that can be checked through algebraic addition of components Fig 36 indicates low frequency...

Страница 28: ...the square wave indicates a transient oscillation in a relatively high Q network in the amplifier circuit In this case the sudden transition in the square wave potential from a sharply rising relativ...

Страница 29: ...tuated fundamental C High frequency loss No phase shift D Low frequency phase shift E Low frequency loss and phase shift F High frequency loss and low frequency phase shift G High frequency loss and p...

Страница 30: ...or EXT TRIG is up to 50V DC AC peak and the input to Z AXIS is up to 50V DC AC peak 4 Do not increase the brightness of the CRT unnecessarily 5 Do not leave the oscilloscope for a long period with a b...

Страница 31: ...adjustments use a well insulated screw driver 3 Before marking adjustments be sure to turn on the power and wait until the unit is stabilized 4 For adjustment follow the procedures described below 5...

Страница 32: ...the trace comes to the position of the line in CH2 NOR mode when the CH2 polarity is inverted Horizontal Position Adjustment 1 To adjust the horizontal position under the nor mal sweep condition set t...

Страница 33: ...SCHEMATIC DIAGRAM CS 1566A 33...

Страница 34: ...A p r o d u c t of T R I O K E N W O O D C O R P O R A T I O N 17 5 2 chome Shibuya Shibuya ku Tokyo 150 Japan 2 5 3 1 2 PRINTED IN J A P A N B50 2918 00 G...

Отзывы: