1) SAFETY
This manual contains information and warnings that must be followed for operating the
instrument safely and maintaining the instrument in a safe operating condition. If the
instrument is used in a manner not specified by the manufacturer, the protection
provided by the instrument may be impaired.
The meter protection rating, against the users, is double insulation per
UL/IEC/EN61010-1 Ed. 3.0, IEC/EN61010-2-033 Ed. 1.0, CAN/CSA C22.2 No. 61010-1
Ed. 3.0, IEC/EN61010-2-032 Ed. 3.0 & IEC/EN61010-031 Ed. 1.1:
Measurement Category III 600V & Category IV 300V AC & DC.
Per IEC61010-1 (2010) OVERVOLTAGE CATEGORY
OVERVOLTAGE CATEGORY ll (CAT II) is for equipment intended to be supplied from
the building wiring. It applies both to plug-connected equipment and to PERMANENTLY
CONNECTED EQUIPMENT.
OVERVOLTAGE CATEGORY lll (CAT III) is for equipment intended to form part of a
building wiring installation. Such equipment includes socket outlets, fuse panels, and
some MAINS installation control equipment.
OVERVOLTAGE CATEGORY lV (CAT IV) is for equipment installed at or near the
origin of the electrical supply to a building, between the building entrance and the main
distribution board. Such equipment may include electricity tariff meters and primary
overcurrent protection devices.
TERMS IN THIS MANUAL
WARNING identifies conditions and actions that could result in serious injury or even
death to the user.
CAUTION identifies conditions and actions that could cause damage or malfunction in
the instrument.
WARNING
To reduce the risk of fire or electric shock, do not expose this product to rain or moisture.
The meter is intended only for indoor use.