4.7
TRICOR Systems Inc.
4.2.5 Plot Baseline Temperature (Option)
The Plot Baseline Temperature is an option that must be purchased when
ordering the Model 225. It is only available as an option if you also purchase the
external printer or TMDAS option. If you do not have these options, the
Plot
Baseline Change
check box in the feature group will be unchecked and the
feature will be unavailable. In addition, the small pushbutton to the right of the
Plot Baseline Temperature
field will be grayed-out.
This menu group allows the baseline of the printed graph to be adjusted. This
option is necessary with some products where the inflection point may occur
above or below the range of the normal printout.
To change the baseline temperature of the external printout, press the small
button next to the
Plot Baseline Temperature
field and enter your new
temperature in the dialog that appears. The range is limited and indicated on
the dialog. Press the
Ok
button to accept the changes and transfer the new
value to the Model 225.
4.2.6 CTU Offset
The CTU offset is a constant added to the CTU calculation. TRICOR uses this
value to adjust the temper meter models so that a consistent reading is obtained
between models. The chocolate manufacturer can also use this value to adjust
for minor differences between temper meters.
The current value of the CTU offset is displayed in the
CTU Offset
field. Press
the small button next to this field to make a change. Enter the new value in the
dialog box that appears. The CTU offset is limited in range to +9.9. Press the
Ok
button to accept the changes and transfer the new value to the Model 225.