Tree MRB 10000g Скачать руководство пользователя страница 13

 

 

13 

Rev 2, 1_19_2016 

NOTE: 

X is the calibration weight which can be set according to your

 

requirements. The weight needed when re-calibrating your Tree scale needs to 
be at least half (1/2) of the capacity of the scale with two thirds (2/3) being ideal 
as this will give you the most accurate reading. 

 

(Ex: For a 10,000g scale you would need at least 5,000g) 

  Press the Zero key to confirm the calibration weight. The display will show CAL 

and then the AD value. Once the 

stable indicator “(.)” is displayed, press the 

UNIT key. The display will flash the calibration weight.

  Place the calibration weight on the platter.  After the stable indicator

“(.)” is displayed, press the UNIT key. The display will show “--------“, then the AD 
value. The calibration is now finished. 

  To return to the weighing mode, press the ON/OFF key to turn the power off. 

Press the ON/OFF key again to turn on the power and return the machine to 
weighing mode. Place the weight on the platter again to ensure the weighing is 
correct. If not, repeat the calibration process.

 

Maintenance & Service

 

 

 

 

 

For  maintenance-work,  the  balance  must  be  disconnected  from  the  power  supply 
(remove  power  adapter  plug  from  socket).  Also  ensure  that  the  balance  cannot  be 
connected to the power supply during the work by a third party. 
 
Make sure that no liquid spills into the scale while performing maintenance work. If 
liquid is spilled on the scale, it must be inspected by a service technician. 
 
Regularly perform maintenance to the weighing pan and the weighing pan holder by 
removing any dirt or dust from under the weighing pan and on the weighing scale 
housing. Use a soft brush or a soft, lint-free cloth, moistened with a mild soap solution

 

 

 

 
Never use solvents, acids, alkalis, paint thinners, scouring powders or other aggressive 
or corrosive chemicals for cleaning; these substances can cause damage to the 
surfaces of the scale housing.

 

 

 
 
 
 
 
 

Содержание MRB 10000g

Страница 1: ...Professional Weighing Equipment MRB Series Mid Resolution Balances Compact Scales MRB 500g MRB 1200g MRB 2500g MRB 6000g MRB 10000g Operation Manual...

Страница 2: ...Customer Service 4 Introduction 4 Safety 5 Weight Scale 6 Functions 7 Scale Description 8 Application Conformity 9 Getting Started 9 Delivery 10 Assembly Installation 10 11 Application Menu 11 12 Cal...

Страница 3: ...ich this declaration refers conforms to the following standards Electronic scale MRB Compact Scales Models MRB 500g MRB 1200g MRB 2500g MRB 6000g MRB 10000g Mark applied EU Directive Standards 2004 10...

Страница 4: ...What you should know about this Operation Manual Tree Professional Weighing Equipment products are simple to operate Nevertheless you should read through these operating instructions in their entiret...

Страница 5: ...adapter If there should be any reason to believe that it is no longer safe to operate the scale it should be immediately unplugged from the electricity supply pull out power adapter and secured again...

Страница 6: ...6 Rev 2 1_19_2016 Weight Scale Construction Functions The weighing scale consists of the following parts The Scale body 1 Scale pan 2 The adapter Operation manual 2 1...

Страница 7: ...oz lb oz 1 2000 1 15000 division available Model number Capacity Division Weighing pan Size MRB 500 500 g 0 1 g 175 x 160 mm MRB 1200 1200 g 0 1 g 175 x 160 mm MRB 2500 2500 g 0 1 g 175 x 160 mm MRB 6...

Страница 8: ...19_2016 Scale Description Weighing platform Battery Compartment Cover Battery Compartment Batteries 6 AA Not included AC adapter socket Weighing Units Stable Indicator Zero Indicator Tare Indicator Lo...

Страница 9: ...pter supplied for the weighing scale complies with the appropriate electrical protection class The following applies to all MRB series weighing machines Power supply Input 120V or 230V AC 15 20 60Hz O...

Страница 10: ...ect delivery for completeness immediately upon unpacking all components Checklist for complete delivery Component delivered present yes no Weighing unit body Weighing pan Power adapter Operating manua...

Страница 11: ...roof horizontal base Make sure that the weighing machine cannot be shaken or knocked over Protect from direct solar radiation Avoid drafts and excessive temperature fluctuations Avoid placing the scal...

Страница 12: ...w seconds and then the 0 will appear Press the UNIT key to select the desired weighing unit kg g lb oz lb oz To begin weighing place the objects to be weighed in a weighing container or directly on th...

Страница 13: ...key again to turn on the power and return the machine to weighing mode Place the weight on the platter again to ensure the weighing is correct If not repeat the calibration process Maintenance Servic...

Страница 14: ...er purchase After 30 days expire the customer is responsible for shipping the product back to us After the product is received we will inspect it and as necessary we will repair or replace and will sh...

Отзывы: