30
31
EMC GUIDANCE
EMC GUIDANCE
EMC Guidance
1)This product needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided,
and this unit can be affected by portable and mobile RF communications
equipment.
2)* Do not use a mobile phone or other devices that emit electromagnetic
fields, near the unit. This may result in incorrect operation of the unit.
3)Caution: This unit has been thoroughly tested and inspected to assure
proper performance and operation!
4)* Caution: This machine should not be used adjacent to or stacked with
other equipment and that if adjacent or stacked use is necessary, this
machine should be observed to verify normal operation in the configuration in
which it will be used.
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
RF emissions
CISPR 11
Group 1
Class B
Class A
Complies
Compliance
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
RF emissions
CISPR 11
Emissions test
Electromagnetic environment - guidance
The device is intended for use in the electromagnetic environment specified below. The customer or the
user of the device should assure that it is used in such an environment.
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
The device is suitable for use in all establishments,
other than domestic and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
±2 kV
power supply lines:
line(s) to line(s):
±1 kV
line(s) to earth:
±2 kV
0%U
T;
0.5 cycle
At 0°, 45°, 90°, 135°,
180°,225°,270° and 315°
0%U
T
; 1 cycle
and
70%U
T
; 25/30 cycles
Single phase: at 0°
0% U
T
; 300 cycle
30 A/m
50Hz/60Hz
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Table 2
Compliance level
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electromagnetic
environment - guidance
30 A/m
50Hz/60Hz
Power frequency
(50Hz/60Hz
)
magnetic field
IEC 61000-4-8
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC61000-4-5
input/output lines:
±1 kV
The device is intended for use in the electromagnetic environment specified below. The customer or the
user of the device should assure that it is used in such an environment.
IEC 60601 test level
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
±2 kV
power supply lines:
100 kHz repetition
frequency
line(s) to line(s):
±1 kV
100 kHz repetition
frequency
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
0% U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°,225°,270° and
315°
0% U
T
; 1 cycle
and
70% U
T
; 25/30 cycles
Single phase: at 0°
0% U
T
;300 cycle