* Be careful to strangulation due to cables and hoses, particularly

  due to excessive length.
* At least 30 min required for ME equipment to warm from the

  minimum storage temperature between uses until it is ready for

  intended use. At least 30 min required for ME equipment to cool

  from the maximum storage temperature between uses until it is

  ready for intended use.
* This equipment needs to be installed and put into service in

   accordance with the information provided in the

   ACCOMPANYING DOCUMENTS;
* Wireless communications equipment such as wireless home

  network devices, mobile phones, cordless telephones and their

  base stations, walkie-talkies can affect this equipment and

  should be kept at least a distance d away from the equipment.

  The distanced is caculated by the MANUFACTURER from

  the 80 MHz to 5.8 GHz column of Table 4 and Table 9 of

  IEC 60601-1-2:2014,as appropriate.
* Please use ACCESSORIES and detachable partes specified/

  authorised by MANUFACTURE.

  Otherwise, it may cause damage to the unit or danger to the

  user/patients.
* There is no luer lock connectors are used in the construction of

   tubing, there is a possibility that they might be inadvertently

   connected to intravascular fluid systems, allowing air to be

   pumped into a blood vessel.
* Please use the device under the environment which was provided in

  the user manual. Otherwise, the performance and lifetime of the

  device will be impacted and reduced.

CAUTION

* This device is intended for adult use in homes only. 
* The device is not suitable for use on neonatal patients, pregnant

   women,patients with implanted, electronical devices, patients with

   pre-eclampsia, premature ventricular beats, atrial fibrillation,

   peripheral, arterial disease and patients undergoing intravascular

   therapy or arterio-venous shunt or people who received a

   mastectomy.

   Please consult your doctor prior to using the unit if you suffer from

   illnesses.
* The device is not suitable for measuring the blood pressure of

   children. Ask your doctor before using it on older children.

28

29

SAFETY INFORMATION

SAFETY INFORMATION

Safety Information

Symbol for “THE OPERATION GUIDE MUST BE READ”  

Symbol for “MANUFACTURER”

Symbol for “SERIAL NUMBER”

Symbol for “TYPE BF APPLIED PARTS” 

Symbol for “DIRECT CURRENT”

Symbol for “ENVIRONMENT PROTECTION - Electrical

waste products should not be disposed of with

household waste. Please recycle where facilities exist.

Check with your local authority or retailer for recycling

advice”

Symbol for “MANUFACTURE DATE” 

Caution: These notes must be observed to prevent any

damage to the device.

SN

The Green Dot is the license symbol of a European

network of industry-funded systems for recycling the

packaging materials of consumer goods.

Symbol for “RECYCLE”

The signs below might be in the user manual,labeling or other

component.

They are the requirement of standard and using.

Содержание LIFEHOOD TMB-1775

Страница 1: ...htly so as to ensure the safe usage of this product and keep the manual well for further reference in case you have problems Guangdong Transtek Medical Electronics Co Ltd Zone B No 105 Dongli Road Tor...

Страница 2: ...Replacing the Batteries DATA MANAGEMENT 17 INFORMATION FOR USER 19 Setting Date Time Contraindications LCD Display Signal General Description Indications for Use Measurement Principle Start the Measur...

Страница 3: ...pulse rate measurement and the result storage The design provides you with two years of reliable service Readings taken by the TMB 1775 are equivalent to those obtained by a trained observer using the...

Страница 4: ...Systolic pressure High blood pressure Diastolic pressure Low blood pressure EXPLANATION mmHg Measurement Unit of the blood pressure Low battery Batteries are low and need to be replaced Current Time Y...

Страница 5: ...22cm 42cm 3 4 AAA batteries Please use LIFEHOOD authorized cuff The size of the actual cuff please refer to the label on the attached cuff Component list of pressure measuring system 1 Cuff 2 Air pipe...

Страница 6: ...ries In order to get the optimal performance and protect your monitor please use the correct batteries CAUTION Do not use new and used batteries together Do not use different types of batteries togeth...

Страница 7: ...Press M button to change the Each press will increase the number by one in a cycling manner When the monitor is off hold pressing START STOP for 3 seconds to enter the mode for year setting 10 11 BEF...

Страница 8: ...your left upper arm Do not place the arm cuff over thick clothes 2 Insert the air plug into the air jack securely 3 Wrap the arm cuff firmly in place around your left upper arm The bottom edge of the...

Страница 9: ...t mmHg mmHg min How to Sit Correctly To take a measurement you need to be relaxed and comfortably seated under comfortable room temperature Avoid bathing drinking alcohol or caffeine smoking exercisin...

Страница 10: ...instead 2 Press the START STOP button to turn the monitor off It will automatically turn off within 1 minute mmHg mmHg min 2 Press the M to get the record you want The current No is No 1 Five records...

Страница 11: ...the selected user by following steps below Delete the Records 1 Hold pressing M button for 3 seconds when the monitor is in the memory recall mode the flash display dEL ALL will show mmHg mmHg min 2...

Страница 12: ...ter clean it with a dry cloth in case Avoid dusty and unstable temperature environment Do not attempt to clean the reusable cuff with water and never immerse the cuff in water What is the standard blo...

Страница 13: ...mend you seek medical advice Please note that the device does not replace a cardiac examination but serves to detect pulse irregularities at an early stage An irregular heartbeat is detected when a he...

Страница 14: ...sure again Relax for a moment and then measure again The measurement process does not detect the pulse signal Loosen the clothing on the arm and then measure again Warning message Relax for a moment R...

Страница 15: ...Mode of operation Continuous operation Degree of protection Protection against ingress of water Software Version A01 Approx 120 2mm 108 2mm 68 5mm 4 AAA batteries user manual IP21 It means the device...

Страница 16: ...d in the user manual Otherwise the performance and lifetime of the device will be impacted and reduced CAUTION This device is intended for adult use in homes only The device is not suitable for use on...

Страница 17: ...n the same limb which other monitoring ME equipment is applied around simultaneously because this could cause temporary loss of function of those simultaneously used monitoring ME equipment On the rar...

Страница 18: ...roper results or serious danger Do not wash the cuff in a washing machine or dishwasher The service life of the cuff may vary by the frequency of washing skin condition and storage state The typical s...

Страница 19: ...meters Part 3 Supplementary requirements for electro mechanical blood pressure measuring systems IEC 80601 2 30 2013 Medical electrical equipment Part 2 30 Particular requirements for the basic safety...

Страница 20: ...d by portable and mobile RF communications equipment 2 Do not use a mobile phone or other devices that emit electromagnetic fields near the unit This may result in incorrect operation of the unit 3 Ca...

Страница 21: ...rcial or hospital environment Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment Powe...

Страница 22: ...itter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is af...

Страница 23: ...c 5kHz deviation1kHzsine GMRS 460 FRS 460 LTE Band 13 17 Pulse modulation b 217Hz Pulse modulation b 18Hz Pulse modulation b 217Hz Pulse modulation b 217Hz Pulse modulation b 217Hz GSM 800 900 TETRA 8...

Страница 24: ...he reduced minimum separation distance Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following equation Where P is the maximum power in W d is the minimum...

Отзывы: