background image

OR

Selecting Colors:
Depress the End Button [B] to cycle 

through these modes:
Orange  Optimal wavelength for highlighting the 

contrast between the venous blood 

vessels and surrounding tissue.

Green  Assist in finding arteries and veins.
White 

Exam light.

Off

Cleaning Directions
Cleaning the device will not be necessary if properly 

used with the disposable plastic cover.
In the event that the device needs to be cleaned, 

use germicidal wipes sparingly.
Do not submerge the unit in liquid. 

Protective Cap:
The purpose of the Protective Cap [E] is to protect 

the End Button [B], Side Button [C], and LED 

Window [A] while the device is not in use.
Keep a disposable plastic cover on device to 

ensure a proper fit.

Attaching an optional Lanyard:
Attach lanyard to the Cap [E

then snap it over the 

End Button [B].
The Protective Cap and lanyard should be cleaned 

regularly with alcohol or germicidal solution.

Installing/Replacing Batteries:
The Veinlite R is powered by a single 

AAA battery.
1  Press firmly on the Battery Cover [D

near the seam and slide it off.

2  Remove the used battery.
3  Install a new AAA battery, 

oriented as indicated by the 

markings inside the battery 

compartment.

4  Slide the Battery Cover into place.

Warranty
The Veinlite R has a one year parts & labor warranty.

Troubleshooting
Before returning the device to TransLite 

for repair, please call or e-mail to obtain an 

authorization number and shipping details.
No refunds will be given for 

devices shipped without 

authorization.

LED Failure
If one of the LED fails to turn on please replace the 

batteries. If the problem persists contact TransLite 

for assistance.

No Light From Any LEDs
Replace the battery, then try turning the unit on 

again. If the device still will not turn on, please 

contact TransLite for assistance.

Specifications
Number of LEDs: 

1 Orange

  

1 Green

  

1 White

Weight with battery: 

29g

Dimensions: 

96.1×19.0×16.5mm

Battery: 

AAA 1.5V

Explanation of Symbols
 

Model Number

 

Serial Number

 Manufacturer

 

EC Representative (EU Only)

 

Do not dispose of as unsorted municipal waste

M. Devices Group

Health Education Centre

The Church, Portland Street

Southport PR8 1HU

United Kingdom

+44 1704 544 944

Operating Directions

Disposable Covers:
The Veinlite R should always be used with a 

single-use disposable plastic cover to protect it from 

contamination and to prevent the spread of 

infection.
1  Place the Veinlite R inside 

the disposable plastic cover. 

Check the cover to make sure it 

is intact and the flap is on the 

side away from the LED 

Window [A].

2  These covers are not sterile. 

They should be wiped 

with a germicidal wipe 

prior to use.

3  Discard the single-use 

disposable plastic cover after each 

subject.

Disposable plastic covers are not included and 

must be ordered fromTransLite or any of its 

authorized dealers.

Device:
1  
Install a new Cover [F].
2  Slide Large Tail Holder [G] or Small Tail 

Holder [H] over the LED Window [A] to 

allow illumination of tail veins.

3  Place Veinlite R against the 

tail with the Tail Holder in 

contact with the tail.

4  Depress the End Button [B] to 

turn on the LED.

5  Apply gentle pressure to the Veinlite R to 

maintain optimal contact.

6  Device will turn off after cycling 

through the colors or you can 

depress and hold the End Button [B] for one 

second to turn device off.

7  When you turn on a device with a low battery the 

device will shut itself off.

8  Device will shut off after approximately 4 minutes.

Brightness Control:
There are 3 brightness levels.
 Depress the 

Side Button [C] to change the brightness. Each 

depression of the button will cycle through low, 

medium, and high brightness levels. The lowest 

brightness level is the default setting.

Caution
Veinlite R is designed for external use only.
To avoid contamination of the device, Veinlite R must 

be used with the disposable plastic cover.
To avoid cross contamination, always change the 

disposable plastic cover between subjects and wipe 

the new cover with a recommended germicidal wipe.
Always turn the LEDs on after the device is in 

position. Do not look directly at the LEDs when they 

are illuminated or shine the light from them directly 

into anyone’s eyes.
Do not immerse the device in liquid.
Do not autoclave the device.

Introduction
The Veinlite R is a hand-held battery powered 

transillumination device for visualizing veins and 

arteries in the tails of rats and mice.

Applications
To assist in venous and arterial access in rats and 

mice tails.

      

READ INSTRUCTIONS BEFORE USE

TransLite LLC

345 Commerce Green Blvd

Sugar Land TX 77478 USA

+1 281 240 3111

www.veinlite.com

[email protected]

LED Window [A]

End Button [B]

Cap [E]

Cover [F]

Large Tail Holder [G]

Small Tail Holder [H]

Side Button [C]

Battery Cover [D]

Veinlite 