Calibration 

 

 

The CrossFLOW's ion output is inherently balanced by design, so there 
are no calibration adjustments. Periodic verification is recommended, 
per ANSI/ESD SP3.3-2006 Periodic Verification of Air Ionizers. Verifi-
cation frequency may be set by the individual user, but is recom-
mended to be not less than one time per year.  
 
Equipment needed is a Charge Plate Monitor such as Transforming 
Technologies 287A or 288B.  Test the ionizer's ion output and balance  
according to the Charge Plate's instructions.   
 
Ion Output Check: Discharge times 

should be 2 seconds (1000v-100v) at 

12".  

Ion Balance Check: Offset

 voltage should be measured and be 10 

volts. 
 

 
 
 

 

 
 
 
 
 
 

287A -Charge Plate

 

288B - Charge Plate

 

Содержание CrossFLOW IN3000

Страница 1: ...Model IN3000 Instruction Manual T R A N S F O R M I N G T E C H N O L O G I E S L L C O U T S T A N D I N G A L T E R N A T I V E S I N S T A T I C C O N T R O L...

Страница 2: ...onized air Inherently balanced ion output Stepless fan speed adjustment for a wide range of air flow Heater for operator comfort Bench stand allows for angled output Built in ion emitter cleaner Ioniz...

Страница 3: ...the ionizer and or personal injury Keep the unit dry Do not operate the ionizer in a flammable volatile or explosive atmosphere Do not insert objects through the unit s intake or outlet grilles Damag...

Страница 4: ...se cure it The CrossFLOW should be placed approximately 1 to 3 feet from the critical work area or objects to be neutralized It should be positioned to cover as much of the area as possible with the i...

Страница 5: ...fast neutralizing the item should be held within 1 to 3 feet from the CrossFLOW For virtually instantaneous neutraliz ing the item may be held as close as 6 inches Maintenance The CrossFLOW has been d...

Страница 6: ...set by the individual user but is recom mended to be not less than one time per year Equipment needed is a Charge Plate Monitor such as Transforming Technologies 287A or 288B Test the ionizer s ion ou...

Страница 7: ...ccident In no event shall Transforming Technologies be liable for collateral or consequential damages To receive service un der warranty please contact Transforming Technologies Technical Support Abou...

Страница 8: ...3719 King Rd Toledo OH 43617 P 419 841 9552 F 419 841 3241 info transforming technologies com...

Отзывы: