CHAPTER 3: CONFIGURATION  

 

3.1 CONFIGURATION 

OVERVIEW 

 

The scale contains two main setup menus: The Setup (“F”) menu, which configures the scale to 
your weigh platform and the User (“A”) menu, which configures the serial communication port and 
enables some user options. The Setup and User menus consist of several menu selections, each with its 
own sub-menu of choices.  
 
To set up the scale, you must first enter the appropriate menu mode. Once there, four of the front panel keys 
become directional navigators to move around in the menus, and one key is used to save or SET the 
selections. 

 

3.2 

SETUP (“F”) MENU 

 

3.2.1 

ENTERING THE SETUP MENU  

 
1.  Power off the scale by toggling the ON/OFF switch. 

 

2.  Press and hold the numeric 0 key while powering the scale back on. When the scale 

shows ” F   1” , you are in Setup Menu mode and you may release the 0 key. 

 
3.2.2  NAVIGATING IN THE SETUP MENU 
 

Use the directional keys shown in Figure 3-1 to move around in the Setup Menu Chart 
shown in Figure 3-2 on the following page. Detailed descriptions of the Setup Menu can be 
found in Chapter 4.

 

 

1. 

To move to a new “F” heading, use the Print (left) or 9 (right) key to move right or left 
in the Setup Menu Chart. 

 
2. 

To move to the selection level, press the 6 (down) key once. The current saved 
selection is shown. 

 
3. 

To view the available selections for the current “F” heading, use the Print (left) or 9 
(right) key to move through the selection field. 

 
4. 

To save a new selection, press the SAMPLE / SAVE (Set) key. To exit without saving, 
press the 3 (up) key to return to the current “F” heading.  

 
5. 

Repeat Steps 2 through 5 until the Setup Menu is programmed.  

 
 

SETUP MENU MODE KEY FUNCTIONS

6

SAMPLE

9

SET

3

PRINT

 

 

Figure 3-1: Setup Menu Key Assignments 

 

Continued => 

 

Page 3-1 

 

Содержание TC-100 Series

Страница 1: ...1 0 January 10 2003 2003 Transcell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60089 Tel 847 419 9180 Fax 847 419 1515 E m...

Страница 2: ...1 3 2 3 Notes on the Setup Menu 3 2 3 2 4 Exiting the Setup Menu 3 2 3 3 User A Menu 3 2 Chapter 4 Setup Menu Descriptions and Procedures 4 1 4 1 Setup Menu Descriptions 4 1 4 2 Resetting Factory Def...

Страница 3: ...menu which configures the Scale to your weigh platform and the User A menu which configures the serial communication port and enables some user op tions Chapter 3 gives an overview and explains how t...

Страница 4: ...ed for connection to either a PC or a serial printer It can also be connected to a simplex device such as a remote display COM1 is wired to a physical connector a D SUB9 female Refer to the TC 100 Ope...

Страница 5: ...e scale shows F 1 you are in Setup Menu mode and you may release the 0 key 3 2 2 NAVIGATING IN THE SETUP MENU Use the directional keys shown in Figure 3 1 to move around in the Setup Menu Chart shown...

Страница 6: ...ETUP MENU 1 Detailed descriptions of the setup menu parameters can be found in Chapter 5 of this manual 3 2 4 EXITING THE SETUP MENU 1 Power off the scale by toggling the ON OFF switch 2 Power on the...

Страница 7: ...nit to be used in the calibration process Also the default unit for normal operation 1 primary unit is lb 2 primary unit is in kg 1 2 depends on model F9 Display Divisions Determines the desired weigh...

Страница 8: ...while switching the scale back on You may release that key when the scale displays the corresponding message found in Table 4 1 MODEL CAPACITY GRADUATION KEY MESSAGE TC 100 2 2 lb x 0 0002 lb 1 INIT 2...

Страница 9: ...Page 5 1 CHAPTER 5 USER MENU DESCRIPTIONS AND PROCEDURES 5 1 USER MENU DESCRIPTIONS Please refer to Chapter 5 of the TC 100 Operation Manual...

Страница 10: ...1 While in the Setup mode scroll to F 17 then scroll down once using the 6 key to enter span calibration menu 2 The display will momentarily show C 1 for the span calibration followed by a value with...

Отзывы: