CPR Guidance
The i-PAD NF1200-T gives guidance in the form of prompts and beat rhythm during CPR. Following is
the description of the guidance sequence.
Step
Description
Your Action(s)
1
Voice Prompt:
Push the chest down fast two
inches
• Place your hands, one on top of another on the
area between the nipples of the patient
• Push the chest down 2 inches
2
Beat sound:
Rhythm beat sound played at 100
beats per minute.
The number of beats depends on
the Compression - Ventilation
ratio setting
30:2 setting: 30 beats
15:2 setting: 15 beats
• Push the chest of the patient hard and fast in
time with the beat provided by the i-PAD
NF1200-T.
3
Voice Prompt:
Give two breaths
• Immediately open the airway using a head tilt-
chin lift maneuver.
4
Voice Prompt:
Breathe, breathe
• Give two breaths
• Each breath must last for 1 second.
• Each breath must have enough volume to
produce visible chest rise
• Give the two breaths within 5 seconds
5
Cycles:
Steps 1 to 4 are repeated for 5
cycles
• Do 5 cycles of CPR.
NF1200-T
22
CU Medical Systems, Inc.