Trainer NF1200-T Скачать руководство пользователя страница 22

CPR Guidance

The i-PAD NF1200-T gives guidance in the form of prompts and beat rhythm during CPR. Following is 
the description of the guidance sequence.

Step

Description

Your Action(s)

1

Voice Prompt: 

Push the chest down fast two 
inches

• Place your hands, one on top of another on the 

   area between the nipples of the patient

• Push the chest down 2 inches

2

Beat sound:

Rhythm beat sound played at 100   
    beats per minute.
The number of beats depends on 
   the Compression - Ventilation
   ratio setting
30:2 setting: 30 beats
15:2 setting: 15 beats

•  Push  the  chest  of  the  patient  hard  and  fast  in

    time  with  the  beat  provided  by  the  i-PAD 
   NF1200-T.

3

Voice Prompt: 

Give two breaths

•  Immediately  open  the  airway  using  a  head  tilt-

    chin lift maneuver.

4

Voice Prompt: 

Breathe, breathe

• Give two breaths
• Each breath must last for 1 second.
• Each  breath  must  have  enough  volume  to 

   produce visible chest rise

• Give the two breaths within 5 seconds

5

Cycles:

Steps 1 to 4 are repeated for 5 
cycles

• Do 5 cycles of CPR.

NF1200-T

22

CU Medical Systems, Inc.

Содержание NF1200-T

Страница 1: ...i PAD NF1200 T Operator s Manual AED Trainer...

Страница 2: ...up the i PAD 8 Package Contents 8 Infrared Remote Controller Remocon 9 Rescue Scenarios 11 Setting up the i PAD NF1200 T 14 4 Using the i PAD NF1200 T 15 Simulated Rescue Preparation 15 Simulated Mai...

Страница 3: ...ended rescue protocol of international rescue policy making bodies such as the American Heart Association the European Resuscitation Council and the American Red Cross It is to be used in conjunction...

Страница 4: ...ntation NF1200 T 4 CU Medical Systems Inc State LED Power ON OFF button Remote control sensor i Button Handle Cover Turn ON First 1 ON OFF Then Open 1 Turn ON 2 Top View of the i PAD NF1200 T with its...

Страница 5: ...AD NF1200 T blinking green the i PAD NF1200 T is in standby mode and ready for a rescue operation solid green the i PAD NF1200 T is in rescue mode blinking red the i PAD NF1200 T detected a system err...

Страница 6: ...g the connector of the AED electrode pads into this port Graphical Rescue Guide Guides you by indicating the current step in the simulated rescue process Quick Reference Guide A printed card that summ...

Страница 7: ...5 non condensing Do not store the device near heating equipment Do not install or store in environments where humidity tends to rapidly increase or there is a ventilation problem Do not store the devi...

Страница 8: ...ntains the following items NF1200 T 8 CU Medical Systems Inc Turn ON First 1 ON OFF Then Open 1 Turn ON 2 i PAD NF1200 T Disposable non rechargeable battery pack Rechargeable battery pack option Batte...

Страница 9: ...roller should be pointed at the remote controller sensor of the i PAD NF1200 T when controlling its operation The instructor should control the i PAD NF1200 T using this Remote Controller only Any cha...

Страница 10: ...trainee attaches the pads to the manikin pressing this button simulates the attachment of the pads and starts rescue scenario 1 LOW BAT Simulates a low battery condition when pressed UP Increases the...

Страница 11: ...Shock delivery c CPR administration 30 chest compressions 2 artificial breaths d Shockable rhythm e Second Shock delivery f Normal Sinus Rhythm g CPR administration 30 chest compressions 2 artificial...

Страница 12: ...y c CPR administration 30 chest compressions 2 artificial breaths d Shockable rhythm e Second shock delivery f CPR administration 30 chest compressions 2 artificial breaths g Normal Sinus Rhythm h CPR...

Страница 13: ...ificial breaths g Shockable rhythm h Third shock delivery i CPR administration 30 chest compressions 2 artificial breaths j Normal Sinus Rhythm k CPR administration 30 chest compressions 2 artificial...

Страница 14: ...NF1200 T Study the functions of the buttons switches indicators and connection ports Place the trainer defibrillator pads package as shown in the following figures Install the battery pack Place the...

Страница 15: ...State LED turns solid blue Beeper emits a short continuous beep State LED turns solid green Voice prompt Attach pads The voice prompt is played until the training instructor presses the PAD ON OFF or...

Страница 16: ...pads 5 Take the pads out of its packaging NF1200 T 16 CU Medical Systems Inc CAUTION To the training instructor Emphasize the following caution to the trainee Do not waste time in removing the patien...

Страница 17: ...the AED electrode pads connector port of the i PAD NF1200 T 7 Look at the graphics on the back of the pads The graphics indicate the positions of the pads on the patient s body 8 Peel off the protect...

Страница 18: ...ted The i PAD NF1200 T begins rhythm analysis simulation as soon as the training instructor presses the PAD ON OFF button or any scenario button from S1 to S8 The i PAD NF1200 T activates the followin...

Страница 19: ...until the i PAD NF1200 T disarms itself if the SHOCK button is not pressed within 15 seconds Voice prompt Press the flashing orange button now Deliver shock now SHOCK button backlight flashing orange...

Страница 20: ...In the real i PAD NF1200 device the CPR protocol depends on the following settings 1 Rescue Provider Lay Rescuer Healthcare Provider 2 Compression Ventilation Ratio 30 2 30 compressions 2 breaths 15 2...

Страница 21: ...has been delivered Voice Prompt Check pulse Beeper 1 second continuous beep i PAD NF1200 T pauses for 10 seconds to allow the trainee to check the pulse of the simulated patient Voice Prompt Begin CP...

Страница 22: ...nute The number of beats depends on the Compression Ventilation ratio setting 30 2 setting 30 beats 15 2 setting 15 beats Push the chest of the patient hard and fast in time with the beat provided by...

Страница 23: ...the result of service or modification by an entity other than CU Medical Systems Inc or its authorized representatives IN NO EVENT SHALL CU MEDICAL SYSTEMS BE LIABLE FOR CONSEQUENTIAL DAMAGES Only pr...

Страница 24: ...d by the user When the i PAD NF1200 T is not operating properly immediately bring it for servicing to an authorized service center Please fill up the following table with the necessary information whe...

Страница 25: ...wha Medical Instrument Complex 1647 1 Dongwha Munmak Wonju Gangwon Republic of Korea Tel 82 33 747 7657 Fax 82 33 747 7659 EU Authorized Respresentative of CU Medical Systems Inc Medical Device Safety...

Страница 26: ...Dongwha Medical Instrument Complex 1647 1 Dongwha Munmak Wonju Gangwon Republic of Korea Tel 82 33 747 7657 Fax 82 33 747 7659 Homepage www cu911 com i PAD NF1200 T AED Trainer...

Отзывы: