Operator’s Manual Ver 1.30
19
WARNING
To the training instructor: Emphasize the following warning to the trainee.
Do not let anybody touch the patient when you press the SHOCK button. Defibrillation shock can
cause operator or bystander injury.
If a shock is needed:
Indicators and Voice prompts:
• Beeper:
1 second continuous beep
• Voice prompt:
“Shock advised”
• Beeper:
beeps continuously until the SHOCK button is pressed or until the i-PAD NF1200-T
disarms itself if the SHOCK button is not pressed within 15 seconds.
• Voice prompt:
“Press the flashing orange button, now. Deliver shock, now”
• SHOCK button backlight:
flashing orange.
You must press the SHOCK button to deliver the simulated shock.
• If you press the SHOCK button, the i-PAD NF1200-T delivers a simulated defibrillating shock. It
informs you of the shock delivery with the voice prompt: “Shock delivered”
• If you do not press the SHOCK button within 15 seconds, the i-PAD NF1200-T disarms itself and
proceeds to CPR Guidance.
If a shock is not needed:
Indicator:
• Voice prompt:
“No shock advised”