User Manual
TRAQC-3 LL
TRAQC-3 LL 12/2018 EN Rev.5
page : 13
5.
Maintenance
Before any maintenance is undertaken the TRAQC-3 LL must be at room
temperature. The sequence heat up or cool down the TRAQC-3 LL to room
temperature is described in chapter 4.3. When room temperature is reached,
turn off the TRAQC-3 LL and unplug the mains power cord.
5.1
Cleaning TRAQC-3 LL heat block well and inserts:
A small amount of dust or pollution can cause jams when between the inserts and the
“heat block well”. To prevent this:
Remove the insert from the “heat block well” at
regular intervals
Clean the “heat block well” with a dry cloth.
If the “heat block well” is very polluted it can be
cleaned with a brass tube wire brush. After cleaning
flush the “heat block well” with compressed air.
Clean the bore and outside of the insert with a dry
cloth.
If the bore of the heat insert is very polluted it can be cleaned with a brass tube
wire brush. The outside with a brass wire brush. After cleaning flush the bore of
the insert with compressed air and clean the outside with a dry cloth.
If the TRAQC-3 LL is not operated for longer periods or moved, remove the insert from
the “heat block well” and store the insert in one of the two “insert storing holes”.
5.2
Cleaning the fan grille:
On top, the TRAQC-3 LL is fitted with an air grille, “upper grill” via which cooling air
enters the TRAQC-3 LL.
Clean the grille at regular intervals (vacuuming or brushing) depending on the level of air
pollution.
5.3
External cleaning:
Clean the outside of the TRAQC-3 LL with a damp cloth and some water or with a solvent
free and mild cleaning agent.