![Tradinco TRAQC-3 LL Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/tradinco/traqc-3-ll/traqc-3-ll_user-manual_1148135014.webp)
User Manual
TRAQC-3 LL
TRAQC-3 LL 12/2018 EN Rev.5
page : 14
5.4
Fuse
With a flat screwdriver (max. 2.5mm) press down both snap fittings of the “fuse drawer”.
Take out the “fuse drawer” and replace both fuses.
Two Fuses: 3.15A antisurge, 230 VAC must be used.
5.5
Storage and/or shipping
Before storage and/or shipping the TRAQC-3 LL must be cooled to a temperature
between 10°C and 30°C. The procedure described in chapter 4.3. When a
temperature between 10°C and 30°C is reached, turn off the TRAQC-3 LL and
unplug the mains power cord.
When the TRAQC-3 LL is shipped / stored:
Place the insulation tube in the “Insulation tube storage hole”.
Remove the insert from the “heat block well” and store the insert in one of the
two “insert storing holes” or in the insert storing hole in the shipping case.
Store it indoors, in a temperature between 0 and 30° Celsius.
Place the case a horizontal position.
5.6
Recalibration
The TRAQC-3 LL is adjusted and tested with measuring equipment in accordance with
recognized national standards prior to delivery.
The TRAQC-3 LL should, depending on the application situation, be inspected at
appropriate intervals. We recommend you to return the calibrator to Tradinco
Instruments at intervals of max. 12 months for recalibration and service.