TrackSense
®
LyoPro
Version 3.3
Page 25 of 53
User Manual
User Manual TrackSense LyoPro – MAN 7
6.1.4.
Disassembling / Assembling Data Loggers
Battery:
Disassemble the battery compartment of the data logger by unscrewing the
top spanner by hand. Then remove the top lid by turning the spanner around
and use it as lift for the lid. Insert the activated* battery according to the
indication on the housing (+ pole downwards). Exchange the o-ring and put the
lid back before tightening the spanner by hand.
*The battery should be placed in separate activator for 60 minutes before
inserted
Thermocouple Sensor:
If not already mounted, connect the thermocouple sensor to the data logger
using the included screwdriver. It is important that the
RED
wire goes into the
RED
screw terminal. And that the
BLUE
wire goes into the
BLUE
terminal. Do
not tighten the screws too hard in order to avoid possible damage of the
thermocouple.