9.
Alternatively, the data loggers can be programmed: Choose
Settings
at the
bottom of the screen. The Logger Settings window will appear. Define
Sample
Rate
and optionally Programmed Start/Stop conditions and activate
Primary
Control
. Choose
Select All
Apply
Close
10.
After the settings are setup, select
Start
. When all selected slots turn green the
data loggers are programmed and will start logging data. Select
Close
11.
Go to
LyoPro
Logger List
. This will showcase all active/programmed data
loggers. Review the details to confirm that they were programmed correctly