Metalyser
®
Deluxe HM2000
Instruction Manual
v2.0
4.5 TROUBLESHOOTING
Whilst your Metalometer
®
is designed to be very reliable, occasional problems
may occur throughout its working life. The following tables are intended to help
you diagnose and resolve these problems simply and quickly. Should you not
be able to resolve the problem please contact your supplier and they will be
able to assist you.
Problem
Probable Cause
Solution
Hi
The metal concentration is
above the maximum limit of
detection.
Dilute the sample.
Lo
The metal concentration is
below the minimum level of
detection
Result higher than
expected
Cell chamber or cell dirty
Clean chamber and
cell
Turbidity present in sample
Filter sample
Result lower than
expected
Reagent not fully dissolved
Shake sample and
check all reagent is
dissolved
Time not elapsed for timed
tests.
Allow time to elapse
Zero sample was turbid,
dirty or coloured.
Clean sample vial and
refresh sample
Содержание METALYSER DELUXE HM2000
Страница 1: ...METALYSER DELUXE HM2000 Portable Heavy Metals Analyser...
Страница 2: ...2 www trace2o com Metalyser Deluxe HM2000 Instruction Manual v2 0 CONTENTS...
Страница 37: ...www trace2o com 37 Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...
Страница 38: ...38 www trace2o com Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...
Страница 39: ...www trace2o com 39 Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...