27
Metalyser
®
Deluxe HM2000
Instruction Manual
v2.0
3.5. INTERFERENCE EFFECTS
The Metalyser
®
has been designed to test very low levels of metals in water
and as such is very sensitive. Due to the interaction of other metals and
organics in the water source, interferences can occur as with any system of
this type.
The Metalyser
®
is designed to function in water sources that might be suitable for
drinking. The buffers chosen mean that in many instances interferences are unlikely to
occur.
3.6. SPECIFICATIONS
Metalyser
®
HM2000 Handheld Unit
Input Power: 4.5
–5.5 V DC, 250mA.
Power Connector Type: mini USB Type B (Bulgin 400 series)
Li-Ion Internal Battery
Output voltage: 3.3V
Output current: 2200mAh
Parameter
Lower Limit
⁺
Upper Limit*
Arsenic (III)
As (III)
5
500
Total Arsenic
As (III+V)
10
500
Copper
Cu
5
500
Mercury
Hg
5
500
Cadmium
Cd
3
500
Lead
Pb
5
500
Zinc
Zn
5
500
Manganese
Mn
5
200
*Upper limit fixed. If the user requires values higher than this then a dilution will be needed.
⁺
Lower limits achievable using the Handset at 60 seconds deposition.
Increased accuracy can be achieved by setting the deposition time to 120 seconds
Variability at 20ppb ± 5% with Metaware software, ± 7.5% with handheld
Accuracy dependent on element measured, sample matrix and type
Содержание METALYSER DELUXE HM2000
Страница 1: ...METALYSER DELUXE HM2000 Portable Heavy Metals Analyser...
Страница 2: ...2 www trace2o com Metalyser Deluxe HM2000 Instruction Manual v2 0 CONTENTS...
Страница 37: ...www trace2o com 37 Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...
Страница 38: ...38 www trace2o com Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...
Страница 39: ...www trace2o com 39 Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...