![Trace2O METALYSER DELUXE HM2000 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/trace2o/metalyser-deluxe-hm2000/metalyser-deluxe-hm2000_instruction-manual_1146884023.webp)
23
Metalyser
®
Deluxe HM2000
Instruction Manual
v2.0
3.3.5. ADDING A STANDARD ADDITION
When prompted, a 20ppb Standard Addition needs to be added to the
SAB. This is achieved by using the pipette (which is set to 280µl). The
20ppb default setting can be changed by pushing the joystick left
and right. Each 140µl increase on the pipette is equivalent to 10ppb.
On the handheld, when prompted to add 20ppb of a particular element(s) then
the highlighted value of 20ppb can be altered by pushing the joystick to the left
and to the right. If the user wants to measure higher levels then a higher value
closer to the concentration level will be required. The accuracy will increase the
closer the Standard
Addition is to the actual concentration. For example, an
unknown sample is expected to be around 15ppb then a 20ppb standard in this
case is sufficient. If the user expects the analysis to be 100ppb then, for example,
a 120ppb standard would be suitable.
How to use a pipette
1. Place a clean pipette tip on the pipette
2. Select the volume required by twisting the dial at the top of the pipette
3. Hold the pipette vertically with the hand wrapped around it and the thumb on
the plunger at the top; press the
button down gently to the first stop on the
top plunger
4. Place the pipette tip in the standard bottle just underneath the top of the
liquid and slowly release the plunger
5. Move the pipette to the SAB and locate the pipette tip through the hole in the
SAB
6. Push the plunger all the way down to the second stop and hold then slowly
withdraw the pipette tip from the SAB
7. To release the pipette tip after use, push the white button at the side of the
pipette
Before carrying out testing, become familiar using the pipette and the two
different stops on the plunger to ensure the correct volume of standard is added.
1. 2. 3. 4. 5. 6. 7.
M
x
Std
Содержание METALYSER DELUXE HM2000
Страница 1: ...METALYSER DELUXE HM2000 Portable Heavy Metals Analyser...
Страница 2: ...2 www trace2o com Metalyser Deluxe HM2000 Instruction Manual v2 0 CONTENTS...
Страница 37: ...www trace2o com 37 Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...
Страница 38: ...38 www trace2o com Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...
Страница 39: ...www trace2o com 39 Metalyser Deluxe HM2000 Instruction Manual v2 0 NOTES...