The instrument and its accessories are designed for operating at temperatures 
between 5°C and 40°C. The storage temperature can be between 5°C and 
60°C. 

 

Temperatures which are above 60°C can easily be reached in sunshine behind 

glass windows (e.g. in the car)! In order to avoid damage through high 
temperature, it is advisable not to keep or store the instrument or accessories in 
such places. 

 

The instrument and accessories (particularly the plug power supply) should not 
come into direct contact with fluids because of the risk of short circuits! The 
instrument or accessories may only be operated, kept or stored in places where 
the relative humidity is between 30% and 90% (not condensing). 

 

Since the instrument and accessories are not acid proof, any direct contact with 
acid or acidic liquids must be avoided! 

 

The instrument and accessories are not suitable for operating in explosion 

endangered surroundings! 
 

1.5.

 

Opening of the instrument POWDERSAVE II and accessories 

Except for the battery, the POWDERSAVE II and its accessories do not contain 
any components which need to be maintained by the user. 

 

Attention:

 The instrument should not be opened under any circumstances there 

is the risk of getting an electric shock! 

 

1.6.

 

Handling of the probe 

The measuring probe has to be handled with care. It must not be dropped or 
mechanically damaged in any way. To keep the wear on the probe as low as 
possible, observe the following: 

 

Do not place the probe on hot or acidic surfaces, do not immerse into fluids! 

 

Store in dry surroundings! 

 

Powder residues on the active sensor area must be avoided.

 (Responsible 

handling and correct storage prolong the life of the probe mechanism). 
 

Never blow out with a direct stream of compressed air! 

Содержание POWDERSAVE II

Страница 1: ...User Manual February 01 2007 Powder thickness measuring gauge POWDERSAVE II Version 1 2007...

Страница 2: ...f the instrument POWDERSAVE II and accessories 5 2 1 Application 5 2 2 Front and back views 6 2 3 Key functions of the instrument POWDERSAVE II 7 2 4 LCD display 8 2 5 LED display on the instrument PO...

Страница 3: ...ecking after polymerizing with a conventional measuring instrument Only accessories recommended by TQC e g probes cables may be attached to the instrument If the gauge shows signs of damage e g caused...

Страница 4: ...contact with acid or acidic liquids must be avoided The instrument and accessories are not suitable for operating in explosion endangered surroundings 1 5 Opening of the instrument POWDERSAVE II and...

Страница 5: ...antee Waivers TQC does not guarantee the following The instrument or accessories when not used as per these instructions Improper handling such as in explosive areas or hot surroundings Ignoring the r...

Страница 6: ...Page 6 of 26 2 2 Front and back views Illustration View of the probe housing Energy Indicator Distance Indicator Measuring button Display Operating button Bush for Battery charger lead...

Страница 7: ...hown on the display 1st press Start Calibration Calibration The calibration is started using this menu item Attention Calibration can only be commenced when the calibration part is placed on the probe...

Страница 8: ...ems u and L The current upper limit will be displayed If a new value needs to be set then this can be set with the or button The selected value is automatically stored It does not need to be saved sep...

Страница 9: ...of LED on the probe indicates the distance of the probe from the object The optimal measuring distance is reached and the measurements can be registered as soon as the green light on the probe comes...

Страница 10: ...esults or the surrounding conditions have changed the instrument should be recalibrated A correct calibration part is essential if correct measurements are to be made To ensure the faultless state of...

Страница 11: ...ter approx 1 2 months Not using of the Accu for a longer time Loading Unloading increases additionally internal resistance An Accu with increased internal resistance cannot be loaded fully any longer...

Страница 12: ...nt by relevant authorities 6 2 Charging the battery 1 Make sure that your local mains voltage conforms with the details on the label on the bottom of the charger before the first usage 2 Connect the c...

Страница 13: ...ed spirit Attention Do not use thinners 8 Fault finding and messages 8 1 Malfunctions Fault Possible Cause Procedure No display Instrument is not switched on Switch on using GO Battery is discharged R...

Страница 14: ...2 Error messages 99ERR During the measurement an error was arising Measure again Repeated occurrence of 99 ERR Calibrate A coating under 30 was attempted to be measured A coating over 110 was attempte...

Страница 15: ...suring distance to object approx 18 mm Measuring spot diameter approx 1 mm 1 32 Measurement range 30 m 110 m 1 2 mils 4 3 mils Measuring uncertainty 5 m 0 2 mils Measurable layers Powder layers on all...

Страница 16: ...s behind the foam material in the top of the case Carrying case 10 Tips and informations Have a look on our homepage www tqc eu 11 Declaration of Conformity Ramseier Technologies Aktiengesellschaft Fl...

Страница 17: ...two times the arrow key to the right note the indicated value old correction factor 1 switch the equipment off press for about 2 3 sec the HOME button switch the equipment on calibrate the equipment...

Страница 18: ...ickness range 50 70 90 Measuring point Eddy current POWDERSAVE II Eddy current POWDERSAVE II 1 2 3 4 5 6 7 8 9 10 11 12 sum average Old correction factor average new correction factor d1 _____________...

Страница 19: ...e necessary for you to enter both the regular menu and the superuser menu The data and the settings which are dependent on a certain parameter set can only be entered if you access the menu or superus...

Страница 20: ...equipment tests Moreover you are given access to your readings which you can further process in other programs Because of the extended performance range the setting possibilities also have been expand...

Страница 21: ...ersions with a different performance range PsMan2 C Performance range You can change attitudes at the POWDERSAVE II as well as look or take at measurement results Scope of delivery Software and user m...

Страница 22: ...Handling of measurement data s Displaying readings The attitude can comfortably be reached over the menu view and Properties Choose the desired ParameterSet In each ParameterSet 100 measuring data wi...

Страница 23: ...on can be for example managed system dependently Readings with more ParameterSet The readings are put down per ParameterSet To change the ParameterSet at the device you have to change into the Superus...

Страница 24: ...ced in the History The data can be exported directly into an available Excel file prepared Excel file is enclosed with the setup program The prepared Excel file contains macros for the further process...

Страница 25: ...rated again with 2 5 load unloading cycles All these characteristics are NiMh Accu specifically and have nothing to do with the way the Akku is used in the Powdersave II Overloading is very harmful fo...

Страница 26: ...ll be integrated This integrator must be initialized from time to time thus set occasionally on Akku fully or Akku empty So far on the end of the loading procedure the Accu meter was on full initializ...

Отзывы: