TQC CureView AB8000 Скачать руководство пользователя страница 26

|   26

Open the cover of the CureView.

Push the spring behind the probe block 
backward.

Push back the base of the probe block at 
the notch and lift the probe block out.

Fix the nut below the controller.

Replace the controller with a new one. 
Take care not to switch the controllers or 
place them at the wrong locations. Do 
not exert force on the cables connecting 
the probe to the controller.

Lift the controller and guide the probe 
block out through the back of the 
CureView.

10

13

11

12

15

14

Reconnect the connectors. Take care 
the TQC-BUS connector(s) is/are placed 
correctly.

Raise the controller holder plate and 
tighten the allen keys (3mm) on each 
side. 

Tilt and slide in the rear cable cover back 
in place. Fix the 2 allen keys to hold it in 
place.

If the cables prevent a smooth 
positioning. Use the allen key to gently 
push the cables down. To allow  to slide 
the cable cover back in place. 

16

17

18

Содержание CureView AB8000

Страница 1: ...ew Gradient Oven AB8000 REV02 Gradient oven with GOC and Ideal Finish control Operating Instructions V1 2 0918 IMPORTANT Before taking this instrument in use we strongly advise you to read this manual...

Страница 2: ......

Страница 3: ...ion 6 Instrument Controls and Functions 7 Instrument Preparations 7 1 Test Panels 7 2 Calibration Interval 7 3 Lamp Maintenance 7 4 Software 8 Operation 8 1 Preparatory Work 8 2 Sample Preparation 8 3...

Страница 4: ...anyone appointed to work with the instrument should pay attention to the general regulations relating to working safety and accident prevention The safety chapter and the warnings in this manual shou...

Страница 5: ...heated area after the instrument has started Possible immediate danger to the life or health of personnel If this guideline is not noted it can lead to severe danger to health up to fatal injury A da...

Страница 6: ...damage should be documented as accurately as possible possibly photographed and reported to the relevant insurers or in the case of sales delivered to customers works to the supplier 3 4 Storage and P...

Страница 7: ...ght 296 mm 11 65 inch Net weight approx 31 kg 68 34 inch 4 5 Basic Unit Power Supply 220 240 V 50 60 Hz single phase split phase Power consumption max 3000 Watt Display None status indication by 2 LED...

Страница 8: ...12 5 Panel Adapter suitable for Holding Panels size 500 mm x 100 mm Spare parts AB8016 Halogen Gold Reflector Infrared Heating Lamp set of 10 pcs AB8020 Lamp Replacement Tool AB8030 Nextrema Glass Tes...

Страница 9: ...eparation of Energy Connections The instrument is equipped with a safety tested mains supply cable and may only be connected to plug sockets with earth connection complying with the safety regulations...

Страница 10: ...Canopy Safety Release Emergency Stop Test Surface Loading Zone Panel Lifter Audio Alarm Control Buttons Thermal Sensors 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 1 2 3 4 5 Cooling Zone Panel Clamp Hea...

Страница 11: ...each measuring and controlling 8 channels TQC advises to have an annual calibration interval Contact TQC or a local representative for information on how to calibrate the TQC CureView 7 4 Lamp mainte...

Страница 12: ...on as described earlier 8 2 Sample preparation Prepare a sample suitable for conducting the tests as required and suitable for the TQC CureView 8 3 Start the instrument Start the instrument following...

Страница 13: ...no paint or other liquids are spilled on the electronics or left in the holes 9 1 Disposal of Materials Disposal of materials used in the operation of the instrument or for auxiliary functions and ex...

Страница 14: ...ed environment is advised having 20 25 C 68 77 F and 40 60 rH Other ranges of temperature and humidity can limit the minimal programmable temperature of the machine For example if the ambient temperat...

Страница 15: ...graph shows a default slope being equal for all set point temperatures All following examples of ramping and heating graphs are generalized and not correlated to a specific substrate or ramp speed unl...

Страница 16: ...e parameters of the test All these parameters are configured in the software setup mode Default temperature Units Specify in which temperature units the GOC shall display its values by default 12 4 Mo...

Страница 17: ...p to the next stage in the setup mode stage enumerator obtain the same result as selecting a stage by mouse The command will switch the program into run mode if the Pre cooling stage was previously se...

Страница 18: ...ture at the heater specified at Start at heater End at heater Specify the heater at which the gradient shall end End temperature Specify the stop temperature at the heater specified at End at heater D...

Страница 19: ...ollowed by a transport to the heating area After the test the process is reverted and the panel returns to the loading area 12 8 Place panel The gradient oven carrier is waiting in the loading area al...

Страница 20: ...fted and the probes are moved upwards in order to free the panel for transport Transport panel The panel is transported from heating to cooling area Cooling The panel is cooled down according to the c...

Страница 21: ...nce is required on an annual basis Due to the temperature resistant materials used for the production only a hand full of components need to be checked and maintained Maintenance of the TQC CureView s...

Страница 22: ...wer supply and USB cable Loosen the 2 allen keys 3mm that hold the panel lifter in place and lower the panel lifter until it is below the glass bed Unscrew frontpanel allen keys 2 5 mm located one on...

Страница 23: ...ore uniform heat distribution these are mounted alternating On the uneven positions on the back side and at the even to the front Before proceeding always wear the supplied cotton gloves to prevent co...

Страница 24: ...it properly aligns to slide it in the last 1 2 cm Replace the front pressure pin bar and fix the 2 allen keys 17 20 18 21 19 22 Move the panel lifter back up Make sure it sits Replace the front cover...

Страница 25: ...s probes 1 8 Unit 3 Heaters probes 17 24 Unit 2 Heaters probes 9 16 Unit 4 Heaters probes 25 32 Turn machine off Slide the cover downward and tilt it out Unscrew the nut 7mm below the controller Disco...

Страница 26: ...connecting the probe to the controller Lift the controller and guide the probe block out through the back of the CureView 10 13 11 12 15 14 Reconnect the connectors Take care the TQC BUS connector s i...

Страница 27: ...in this manual or otherwise is correct we have no con trol over either the quality or condition of the product or the many factors affecting the use and application of the product Therefore unless we...

Страница 28: ......

Страница 29: ...29 ANNEX A OPERATOR QUALIFICATION LIST Date Name Signature Signature Instructed by Supervisor list allowed to perform lamp replacement and such...

Страница 30: ...30 Date Name Signature Signature Instructed by Operator list...

Страница 31: ...31 ANNEX B MAINTENANCE LIST Date Action Name Signature Installed at test location First run performed...

Страница 32: ...D 40721 Duitsland 49 0 2103 25326 0 49 0 2103 25326 29 info de tqc eu www tqc eu TQC USA Inc 4053 S Lapeer Road Suite H P O Box 108 Metamora Michigan 48455 USA 1 810 678 2400 1 810 678 2422 joel tqc...

Отзывы: