![TQC CureView AB8000 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/tqc/cureview-ab8000/cureview-ab8000_operating-instructions-manual_1146598024.webp)
| 24
Place the middle deflector.
Make sure the front of the glass bed in
inline with the aluminum housing.
Gently tighten the 2 rear allen keys that
hold the glass bed in place.
Slide the glass bed gently back in place.
Make sure it properly aligns to slide it in the
last 1-2 cm.
Replace the front pressure pin bar, and fix
the 2 allen keys.
17
20
18
21
19
22
Move the panel lifter back up. Make sure
it sits.
Replace the front cover and slide it
upwards to fix it in place.
Connect power supply and USB cable.
Verify with a test run that all lamps are
functioning again.
Fix frontpanel allen keys (2.5 mm) located
one on each side to release front cover.
23
24
25
Содержание CureView AB8000
Страница 2: ......
Страница 28: ......
Страница 30: ...30 Date Name Signature Signature Instructed by Operator list...
Страница 31: ...31 ANNEX B MAINTENANCE LIST Date Action Name Signature Installed at test location First run performed...