3. Simulating Millivolt Values
1. Obtain the correct pH / mV output cable for the instrument under test.
2. Connect one end of the pH / mV output cable to the
WP-Sim
and the
other end to the instrument under test, as per the example diagram
below…
3. Switch the
WP-Sim
Function switch to ON. Switch the instrument
under test on and select Millivolt mode. See the instrument’s
handbook if necessary.
4. Switch the pH / mV selector to 650 mV. The readout of the instrument
under test should be approximately 650 mV. Variations are possible
due to the current calibration settings of the instrument under test.