Page 13
4. Switch the pH / mV selector to your choice of pH or mV value. The
readout of the instrument under test should approximately
correspond to the simulated values selected on the
WP-Sim
. In
addition, the pool controller will display the current temperature inside
the
WP-Sim
. Variations are possible due to the current calibration
settings of the instrument under test.
Note : As the
WP-Sim
is simultaneously outputting pH and mV
values, you will notice that BOTH readings on the display change
when the pH / mV selector is changed. Please disregard the pH
values when the
WP-Sim
is set to 650 mV, and disregard the mV
reading when the
WP-Sim
is set to a pH value.