15.
Initialising the WP-84
If the calibration settings of the
WP-84
exceed the allowable limits, the
unit may need to be initialised to factory default values. This action may
be required if the sensor is replaced.
To initialise the
WP-84
…
1. Switch the
WP-84
off.
2. Press and hold the
key while switching the
WP-84
back on.
3. The following messages should be displayed…
Initialized
MUST ReCalibrate
then…
WP84s V4.0 S1234
Con TDS Sal Temp
(The “s” after
WP-84
is shown when the RS232 serial port option is
fitted)
4. The meter then displays Conductivity and Temperature. Note that the
decimal points have been replaced with a “
�
” to indicate that the unit
requires re-calibration.
Note
When the
WP-84
is initialised, the manual k factor selection is re-set to
k=10. See section 14 if you wish to select a k=0.1 sensor.
16.
Instrument firmware version number.
If you need to email, phone or fax TPS for any further technical
assistance, the version number of your
WP-84
firmware may of benefit to
us. The version number is displayed by the
WP-84
at turn-on.