Calibration:
Conductivity/TDS/Salinity Auto
Standard
Recognition
150uS/cm,
1413uS/cm, 2.76mS/cm, 12.88mS/cm,
58.0mS/cm
Conductivity Span Range
k=0.1 Sensor: k=0.075 to k=0.133
k=1.0 Sensor: k=0.75 to k=1.33
k=10 Sensor: k=7.5
to k=13.3
Temperature
Use a reference thermometer.
Temperature Sensor Offset
-10.0
o
C to +10.0
o
C
Memory
3600 readings including date and time.
Automatic Data logging
User-set for one reading every 1 to 90
seconds, 1 to 90 minutes, or 1 to 24
hours.
RS232 Output 300, 1200, 9600 & 19200 baud.
8 bits, no parity, 1 stop bit, XON/XOFF
Protocol.
Clock Calendar clock displays date, month, hours, minutes & seconds.
Year
is
recorded
in
memory
and
transmitted to the RS232 port, but is not
displayed.
Battery Saver On
:
Auto switch-off after 5 minutes.
Off : Continuous use.
Bar Graph display of battery charge
level.
Readout
of
battery
voltage
available for troubleshooting.
Good Laboratory Practices
Date, Time and Value of last Conductivity
and Temperature calibration are stored,
and can be recalled or sent to the
RS232 port at any time.
Power 6V NiMH Rechargeable Battery for approximately 40 hours
operation.
Dimensions
195 x 110 x 55 mm
Mass Instrument only
:
Approx 520g
Full Kit
:
Approx 2.5kg
Environment
Temperature :
0 to 45
O
C
Humidity :
0 to 95 % R.H.