5. Temperature Calibration
5.1 Calibration Procedure
1. Plug the Dissolved Oxygen/Temperature sensor into the
Sensor
socket.
2. Switch the meter on.
3. Place the sensor into a beaker of room temperature water, alongside
a good quality mercury thermometer. Stir the sensor and the
thermometer gently to ensure an even temperature throughout the
beaker.
4. Select Temperature Calibration. (
F1:Cal.
F2:Temp
).
5. The reading from the sensor is now displayed on the right of the
display, and the value you are going to set is shown on the left.
> 25.0< 24*0
o
c
:Set F1:Cal.
6. When the reading on the right has stabilised, press the
or
keys until the reading on the left shows the same temperature as the
mercury thermometer.
7. Press the
key to calibrate the temperature readout.
The
“
” will now be replaced by a decimal point, if calibration was
successful.
Alternatively, press the
key to abort temperature calibration.