Page 23
9. Initialising the 901-C
If the calibration settings of the
901-C
exceed the allowable limits, the unit may
need to be initialised to factory default values. This action may be required if the
sensor is replaced, or when the unit repeatedly fails to calibrate.
To initialise the
901-C
…
1.
Switch the
901-C
OFF
and wait for 10 seconds.
2.
Press and hold the
key while switching the
901-C
back on.
3.
The following messages are now displayed…
Initializing
Initializing
then:
901Cs V1 R1234
901Cs V1 R1234
(The "
s
s
" after "
901C
901C
" is shown when the optional RS232 port is fitted.)
4.
The
901-C
now goes on to Conductivity plus Temperature mode. Note that a
"
∗∗
" replaces each of the decimal points in the Conductivity, TDS and
Temperature readings, indicating that the unit requires calibration.
Notes
When the optional RS232 port is fitted, the Baud Rate is set to 9600 and the Send
Rate is set to zero. See sections 6.1 and 6.2 for details if these settings need to be
altered.
If a k=0.1 Conductivity/TDS sensor is being used, the manual k factor selection
must be re-set to k=0.1 before use.