Page 19
6.6 Data Format
A.
Data is returned to the RS232 port by the
901-C
in the following
format when requested by a PC with the ?D command (section 6.5):
CCCCCCUUU
ss
TTTTTToC
ss
<cr>
or B.
Data is sent to the RS232 port by the
901-C
in the following format
when it is sent by the
901-C
using the Send function (section 6.2):
CCCCCCUUU
ss
TTTTTToC
ss
<cr><lf>
where:
CCCCCC
is the Conductivity or TDS data. maximum 6 characters,
right justified. A “
∗∗
” is sent instead of the decimal point
if the reading is not calibrated.
UUU
is the unit description, left justified. Either “
uS
ss
”, “
mS
ss
”,
“
ppM
” or “
ppK
”. (“
ss
”
is one space.)
ss
is one space.
TTTTTT
is the Temperature data. Maximum 6 characters, right
justified. A “
∗∗
” is sent instead of the decimal point if the
reading is not calibrated.
oC
ss
is the Temperature unit description, left justified. (“
ss
”
is
one space.)
Notes:
1.
+OVR
or
–OVR
is sent when the Data is over-range.
2.
BUSY
<cr> is sent when the
901-C
is Busy (ie in calibration, Baud Rate mode,
Send Rate mode etc.) or when data is not available.