8. TIMED LOGGING
Press and hold down the
Store/Logger Key
for approximately 2 seconds
until the instrument beeps an the following screen is displayed:-
‘S-L9’
will be flashing.
Press the
Down Arrow Key
so that
‘Lo99’
is
flashing then press the
Scroll/Enter Key
and follow
the steps below:-
1.
‘S-t’
will be displayed with
’05 S’
flashing indicating that the instrument
will log a set of 100 readings every 5 seconds.
You can choose from the following logging times
‘05 S’ (100 readings every 5 seconds)
‘10 S’ (100 readings every 10 seconds)
‘20 S’ (100 readings every 20 seconds)
‘30 S’ (100 readings every 30 seconds)
‘01 M’ (100 readings every 1 minute)
‘02 M’ (100 readings every 2 minutes)
‘03 M’ (100 readings every 3 minutes)
by pressing either the Up or Down Arrow Keys
2.
Once you have chosen your required logging period press
the
Scroll/Enter Key
once and
‘PA9E’
will be
displayed.
A page number from 0 to 3 will be flashing.
Select the required page location that you wish to log
the data to by pressing the Up or Down Arrow
Keys
3. Press
the
Scroll/Enter Key
once and the instrument will return to
the previous screen/function with
‘Logger’
flashing on the top line.
The instrument will continue logging until all 100 sets of readings
have been saved or you press and hold down the
Store/Logger Key.
The data logged during this process can ONLY be downloaded to a compatible
PC.
WARNING: -
The instrument will
NOT
turn off if data logging is commencing.
A beep will be heard to warn you of this fact if you try to turn the instrument off.