6. REVIEWING DATA
1. Press
the
Recall Key
once
and the following screen will be
displayed.
‘Stor’
will be flashing on the display. If you wish to review
the Tightness Test Readings or the Last Time Calibrated
Date then press the
Down Arrow Key
to have
‘tItn’
or
‘CAL’
flashing rather than
‘Stor’
2.
Press the
Scroll/Enter Key
once
and
if you have chosen to view
the
Last Time Calibrated
(CAL)
then the Date that the Last Time the 712
was calibrated will be displayed flashing on the screen. (Go To Step 6)
3.
However
if you have chosen to view either Stored
(Stor)
or Tightness Test
(tItn)
Readings then
‘Addr’
will be
displayed and a location number will be flashing.
Select the required address location that you wish to
review by pressing the Up and Down Arrow
Keys.
4. Press
the
Scroll/Enter Key
once and
the
Time & Date of the
Saved Data
from the selected address location will be displayed flashing
on the screen.
**
If there is no data present at that location then
‘nULL dAtA’
will be
displayed flashing**
5. The rest of the Saved Data at this address location can be reviewed
by pressing the Up and Down Arrow Keys
6.
Press the
Scroll/Enter Key
once and
“End YES”
with
“YES”
flashing will be displayed. Should you not
wish to view any further stored data then press the
Scroll/Enter Key
once to Exit and be returned to
the screen you were originally on.
7. However if you wish to continue and review another set
of readings then press the
Up Arrow Key
so that
“no”
is flashing.
8.
Press the
Scroll/Enter Key
once and repeat from Step 2