25
24
7.
Read the capacitor value on the LCD.
i. Measuring Frequency
WARNING!
Never attempt a frequency measurement with a
voltage source greater than 600V. Determine the
voltage of any unknown frequency source before
connecting the instrument in frequency mode.
Instrument set-up:
FUNCTION
BLACK
RED
MINIMUM
MAXIMUM
TEST LEAD
TEST LEAD
READING
READING
Hz
COM
V
Ω
Hz
0.01Hz
200.00KHz
Measurement Procedure:
1.
Disconnect power to circuit to be measured.
2.
Plug black test lead into the
COM
input jack.
3.
Plug red test lead into the
V
Ω
Hz
input jack.
4.
Set the rotary switch to the
Hz
function.
5.
Reconnect power to circuit to be measured.
6.
Read the frequency on the LCD.
h. Measuring Capacitance
WARNING!
All capacitance measurements are to be made
on de-energized circuits with all capacitors
discharged only. Failure to de-energize and dis-
charge capacitors before attempting to measure
them could result in instrument damage and/or
personal injury.
Instrument set-up:
FUNCTION
BLACK
RED
MINIMUM
MAXIMUM
TEST LEAD
TEST LEAD
READING
READING
COM
V
Ω
Hz
0.01µF
10000µF
Measurement Procedure:
1.
Disconnect power to circuit to be measured.
2.
Remove capacitor from the circuit and
discharge it.
3.
Plug black test lead into the
COM
input jack.
4.
Plug the red test lead into the
V
Ω
Hz
input
jack.
5.
Set the rotary switch to the
function.
6.
Connect test leads to capacitor to be
measured.