7. Adjustments
1)
Plug in the drying cabinet and let it run one to two days before storing any items.
2)
Power ON/OFF
•
Press the “ON/OFF” button, internal humidity level will be shown on the LED DISPLAY.
•
During the dehumidification process (for details, please refer to P.2“Dehumidification Process”),
only one “RECYCLE LIGHT” will be on. This light will switch off when the dehumidification process
finishes.
•
To switch off the drying cabinet, press the “ON/OFF” button again.
3)
Adjusting the humidity level
•
Press the “SET” button.
•
Adjust the humidity level by pressing either “UP” or “DOWN” button.
(Initial humidity setting is 1%RH).
•
Press the “SET” button when finish setting. Current humidity level inside the drying cabinet will
be shown on the display.
4)
Display of the temperature level
•
Press the “H/T” button.
•
The “TEMPERATURE DISPLAY LIGHT” will be on. Current internal temperature will be shown on
the LED DISPLAY.
•
This “TEMPERATURE DISPLAY LIGHT” will switch off after 5 seconds automatically and the dis-
play will show the current internal humidity level.
5)
Setting the humidity alarm
(The “HUMIDITY ALARM LIGHT” will be on if the internal RH level exceeds the setting point.)
•
Press both “DOWN” and “SET” button at the same time to set the humidity alarm function.
•
When the display is flashing, adjust the humidity setting point by pressing either “UP” or “DOWN”
button (There is no initial setting, the setting point descends from “99” to “98 by pressing the
“DOWN” button)
•
Press the “SET” button again, the digital and the dot on the right hand side corner will flash.
This digit shows the time for the humidity alarm to activate.
•
Adjust this activated time by pressing either “UP” or “DOWN” button.
•
Press the “SET” button when finish setting, the internal humidity level will be shown on the dis-
play again.
E.g. this light will be on if the internal RH level exceeds 50% for 20 minutes, if the RH setting
point is 50%RH and the time setting value is 20 minutes.
6)
Setting the Door Alarm Buzzer (provided as option, please refer to P.6)