MENU: Calibrate
To re-calibrate the unit based on the center of the plate or to calibrate relative to another
location, follow the steps below in Figure 11. To replace calibration values with those
obtained at the factory prior to shipment, follow steps shown in Figure 12.
NOTE: THE PLATE SURFACE AND BLOCK SURFACE MEASUREMENTS ARE
DIFFICULT TEMPERATURE MEASUREMENT TO MAKE ACCURATELY. Ensure that
the unit and the measurement device have had sufficient time (>10 mins) to stabilize at the
new calibration temperature prior to entering in Measured Temperature Value. If the
temperatures are not stable when the measured temperature value is entered, significant
display and unit performance error may result. Check with the factory if help is needed.
Also, calibration kits are available.
Figure 11:
Menu Map to Calibrate the Front Plate Temperatures