25
−
2-3-1 SR-52 Disassembly and assembly procedure
Illustration
Disassembly procedure
(1) Removal of cover
Remove the 3 screws CR3×8S SUS and then
remove the cover (1004).
(2) Removal of the quick return mirror holder unit
Remove the cover (1004).
Remove the E ring.
Remove the connecting plate (1017) from the
shaft.
Remove the 4 screws CR2×3S.
Remove the quick return mirror holder unit.
Cover (1004)
CR3×8S SUS:3 pcs.
1
(1017)
(EJ2)
E ring
Connecting
plate
2
3
CR2×3S:4 pcs.
Quick return mirror
holder unit
4
1
1
2
3
4
5